Suppr超能文献

三种常见调强放疗递送技术的分次内和分次间可重复性。

The inter- and intrafraction reproducibilities of three common IMRT delivery techniques.

机构信息

Department of Radiology, The University of Texas Health Science Center at San Antonio, San Antonio, Texas 78229, USA.

出版信息

Med Phys. 2010 Sep;37(9):4854-60. doi: 10.1118/1.3476413.

Abstract

PURPOSE

Intensity modulated radiation therapy (IMRT) treatment delivery requires higher precision than conventional 3D treatment delivery because of the sensitivity of the resulting dose to small geometric misalignment of the modulated beamlets. The chosen treatment delivery technique will affect the treatment precision in different ways, based on the characteristics of the delivery method. Delivery using a multileaf collimator (MLC) can reduce treatment time and therapist workload, but typically requires a greater number of monitor units and the fields are prone to both systematic and random leaf positioning errors. An alternative to MLC-based fields, patient specific brass compensators, do not suffer from these leaf positioning errors. In our study, we set out to investigate which delivery method will provide the highest levels of dosimetric reproducibility and the minimum amount of interfraction variability.

METHODS

In our study, a seven field IMRT plan for a head and neck treatment was created using the Pinnacle3 treatment planning system and the intensity maps for each field were obtained. The intensity maps of the fields were delivered with a Varian 2100C/D linear accelerator, using solid compensators and sliding window (SW) and step-and-shoot (SS) MLC segments. Three fields were selected from the seven-beam IMRT plan for comparison. Analysis was carried out using the MatriXX ion chamber array, radiochromic film, and Varian dynalog files.

RESULTS

Our results show that the error in MLC leaf positioning has no gantry angle dependence. The compensator and SW deliveries showed excellent agreement, even when stricter than usual gamma criteria were applied. However, we noted that under these strict conditions, the SS field had at least ten times more pixels out of range than did the compensators. When using step-and-shoot MLC fields, it was observed that the increase in dose rate or the increase of MU/segment degrades the quality of the plan. Analysis of the dynalog files showed that while each individual field had its own propensity for error, all fields showed the same trend: a greater percentage of time the leaves are out of position as dose rate increases, MUs decrease, or both.

CONCLUSIONS

The compensator-based field and both types of MLC-based fields have MatriXX results that are within the clinically acceptable tolerance of 3% dose difference and 2 mm DTA. However, when the criteria are tightened, it becomes evident that the compensators have a definite advantage over their comparable MLC-based competitors in terms of interfraction reproducibility. Fewer monitor units are required to deliver each portal, potentially improving patient outcomes and reducing unwanted side effects to both patients and therapists. In centers without MLC, compensators represent a simple and cost effective way to offer patients state of the art treatment. Based on the results of this study, compensator-based IMRT is a reliable, viable option for use in clinics both with and without MLC-equipped linacs.

摘要

目的

调强放射治疗(IMRT)治疗需要比传统的 3D 治疗更高的精度,因为调制射束的剂量对调制射束的微小几何对准误差很敏感。所选择的治疗输送技术将根据输送方法的特点,以不同的方式影响治疗精度。使用多叶准直器(MLC)进行输送可以减少治疗时间和治疗师的工作量,但通常需要更多的监测单位,并且射野容易出现系统性和随机性叶片定位误差。MLC 叶片定位误差的替代方法是患者专用的黄铜补偿器,它不会出现这些叶片定位误差。在我们的研究中,我们旨在研究哪种输送方法将提供最高水平的剂量学可重复性和最小的分次间变异性。

方法

在我们的研究中,使用 Pinnacle3 治疗计划系统为头颈部治疗创建了七个射野的调强放射治疗计划,并获得了每个射野的强度图。使用瓦里安 2100C/D 直线加速器,使用实心补偿器和滑动窗口(SW)和步进和射击(SS)MLC 段来输送强度图。从七射野调强放射治疗计划中选择三个射野进行比较。使用 MatriXX 离子室阵列、光致变色胶片和瓦里安 dynalog 文件进行分析。

结果

我们的结果表明,MLC 叶片定位误差与机架角度无关。补偿器和 SW 输送显示出极好的一致性,即使应用了比通常更严格的伽马标准也是如此。然而,我们注意到,在这些严格的条件下,SS 场的出界像素数至少是补偿器的十倍。当使用步进和射击 MLC 场时,观察到剂量率的增加或 MU/段的增加会降低计划的质量。对 dynalog 文件的分析表明,尽管每个单独的场都有其自身的误差倾向,但所有场都显示出相同的趋势:随着剂量率的增加、MU 减少或两者兼而有之,叶片位置不正确的时间百分比增加。

结论

基于补偿器的射野和两种类型的基于 MLC 的射野的 MatriXX 结果都在临床可接受的 3%剂量差异和 2mm DTA 容限内。然而,当标准收紧时,补偿器在分次间可重复性方面明显优于其可比的基于 MLC 的竞争对手。每个射野输送所需的监测单位更少,这可能会改善患者的预后,并减少患者和治疗师的不必要的副作用。在没有 MLC 的中心,补偿器代表了一种简单且具有成本效益的方法,可为患者提供最先进的治疗。基于这项研究的结果,基于补偿器的调强放射治疗是一种可靠、可行的选择,可用于配备和不配备 MLC 的直线加速器的临床治疗。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验