Bhuta A C, Babu J P, Jindal G D, Parulkar G B
Electronics Division, Bhabha Atomic Research Centre, Bombay, Maharashtra.
J Postgrad Med. 1990 Apr;36(2):64-70.
This paper describes the basic methods for measurement of body impedance, electrodes and their configuration, and the measuring instrument with its limitations. A microcomputer assisted impedance plethysmograph system, developed at BARC and different lead configurations for impedance plethysmographic investigation are also described. Typical impedance plethysmographic waveforms recorded from a normal subject and measurement of their amplitude and various time intervals are illustrated.
本文描述了测量人体阻抗的基本方法、电极及其配置,以及测量仪器及其局限性。还介绍了在巴巴原子研究中心开发的微机辅助阻抗体积描记系统,以及用于阻抗体积描记研究的不同导联配置。展示了从正常受试者记录的典型阻抗体积描记波形及其幅度和各种时间间隔的测量结果。