Standard design for National Ignition Facility x-ray streak and framing cameras.

作者信息

Kimbrough J R, Bell P M, Bradley D K, Holder J P, Kalantar D K, MacPhee A G, Telford S

机构信息

Lawrence Livermore National Laboratory, Livermore, California 94551-0808, USA.

出版信息

Rev Sci Instrum. 2010 Oct;81(10):10E530. doi: 10.1063/1.3496990.

Abstract

The x-ray streak camera and x-ray framing camera for the National Ignition Facility were redesigned to improve electromagnetic pulse hardening, protect high voltage circuits from pressure transients, and maximize the use of common parts and operational software. Both instruments use the same PC104 based controller, interface, power supply, charge coupled device camera, protective hermetically sealed housing, and mechanical interfaces. Communication is over fiber optics with identical facility hardware for both instruments. Each has three triggers that can be either fiber optic or coax. High voltage protection consists of a vacuum sensor to enable the high voltage and pulsed microchannel plate phosphor voltage. In the streak camera, the high voltage is removed after the sweep. Both rely on the hardened aluminum box and a custom power supply to reduce electromagnetic pulse/electromagnetic interference (EMP/EMI) getting into the electronics. In addition, the streak camera has an EMP/EMI shield enclosing the front of the streak tube.

摘要

国家点火装置的X射线条纹相机和X射线分幅相机进行了重新设计,以增强电磁脉冲硬化能力,保护高压电路免受压力瞬变影响,并最大限度地利用通用部件和操作软件。这两种仪器都使用相同的基于PC104的控制器、接口、电源、电荷耦合器件相机、保护性气密外壳和机械接口。通过光纤与两种仪器相同的设施硬件进行通信。每种仪器都有三个触发器,既可以是光纤的,也可以是同轴的。高压保护包括一个真空传感器,用于启用高压和脉冲微通道板磷光体电压。在条纹相机中,扫描结束后会移除高压。两者都依靠硬化铝箱和定制电源来减少进入电子设备的电磁脉冲/电磁干扰(EMP/EMI)。此外,条纹相机还有一个EMP/EMI屏蔽罩,包围着条纹管的前端。

相似文献

2
Note: X-ray streak camera sweep speed calibration.
Rev Sci Instrum. 2010 May;81(5):056108. doi: 10.1063/1.3429939.
3
Ultra fast x-ray streak camera for ten inch manipulator based platforms.
Rev Sci Instrum. 2012 Oct;83(10):10E106. doi: 10.1063/1.4729500.
8
Experimental study of neutron induced background noise on gated x-ray framing cameras.
Rev Sci Instrum. 2010 Oct;81(10):10E515. doi: 10.1063/1.3478636.
10
Electronic measurement of microchannel plate pulse height distributions.
Rev Sci Instrum. 2010 Oct;81(10):10E310. doi: 10.1063/1.3478687.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验