新型血管内除颤器:在犬模型中比较血管内除颤器与传统植入式心脏复律除颤器的除颤阈值。

Novel intravascular defibrillator: defibrillation thresholds of intravascular cardioverter-defibrillator compared to conventional implantable cardioverter-defibrillator in a canine model.

机构信息

InnerPulse, Inc., Research Triangle Park, North Carolina, USA.

出版信息

Heart Rhythm. 2011 Feb;8(2):288-92. doi: 10.1016/j.hrthm.2010.10.031. Epub 2010 Oct 27.

Abstract

BACKGROUND

An intravascular, percutaneously placed implantable defibrillator (InnerPulse percutaneous intravascular cardioverter-defibrillator [PICD]) with a right ventricular (RV) single-coil lead and titanium electrodes in the superior vena cava (SVC) and the inferior vena cava (IVC) has been developed.

OBJECTIVE

The purpose of this study was to compare defibrillation thresholds (DFTs) of the PICD to those of a conventional implantable cardioverter-defibrillator (ICD) in canines.

METHODS

Eight Bluetick hounds were randomized to initial placement of either a PICD or a conventional ICD. For PICD DFTs, a single-coil RV defibrillator lead was placed in the RV apex, and the device was positioned in the venous vasculature with electrodes in the SVC and IVC. With the conventional ICD, an RV lead was placed in the RV apex and an SVC coil was appropriately positioned. The ICD active can (AC) was implanted in a subcutaneous pocket formed in the left anterior chest wall and connected to the lead system. DFT was determined by a three-reversal, step up-down method to estimate the 80% success level. Two configurations were tested for the conventional ICD (#1: RV to SVC+AC; #2: RV to AC). A single configuration (RV to SVC+IVC) was evaluated for the PICD.

RESULTS

Mean PICD DFT was 14.8 ± 1.53 (SE) J. Conventional #1 configuration demonstrated mean DFT of 20.2 ± 2.45 J and #2 of 27.5 ± 1.95 J. The PICD had a significantly lower DFT than the better conventional ICD configuration (#1; mean difference 5.4 ± 2.1 J, P <.05, paired t-test, N = 8).

CONCLUSION

The new intravascular defibrillator had a significantly lower DFT than the conventional ICD in this canine model.

摘要

背景

一种经静脉、经皮植入的可植入除颤器(InnerPulse 经皮血管内除颤器 [PICD]),带有一个右心室(RV)单线圈导联和钛制电极,位于上腔静脉(SVC)和下腔静脉(IVC)中。

目的

本研究旨在比较 PICD 与犬体内常规植入式心脏复律除颤器(ICD)的除颤阈值(DFT)。

方法

8 只 Bluetick 猎犬随机接受 PICD 或常规 ICD 的初始植入。对于 PICD 的 DFT,将一个 RV 单线圈除颤器导联放置在 RV 心尖,并用 SVC 和 IVC 中的电极将设备定位在静脉脉管系统中。对于常规 ICD,将一个 RV 导联放置在 RV 心尖,将 SVC 线圈放置在适当位置。将 ICD 主动罐(AC)植入在左前胸壁形成的皮下袋中,并与导联系统相连。DFT 通过三反转、逐步上升-下降法确定,以估计 80%的成功率。对常规 ICD 进行了两种配置测试(#1:RV 到 SVC+AC;#2:RV 到 AC)。对 PICD 进行了一种配置(RV 到 SVC+IVC)的评估。

结果

PICD 的平均 DFT 为 14.8 ± 1.53(SE)J。常规#1 配置的平均 DFT 为 20.2 ± 2.45 J,#2 为 27.5 ± 1.95 J。PICD 的 DFT 明显低于较好的常规 ICD 配置(#1;平均差异 5.4 ± 2.1 J,P<.05,配对 t 检验,N = 8)。

结论

在该犬模型中,新型经静脉除颤器的 DFT 明显低于常规 ICD。