Heart Rhythm. 2014 Feb;11(2):210-5. doi: 10.1016/j.hrthm.2013.10.035. Epub 2013 Oct 18.
A percutaneous intravascular cardioverter-defibrillator (PICD) has been developed with a right ventricular (RV) single-coil lead and titanium electrodes in the superior vena cava (SVC)-brachiocephalic vein (BCV) region and the inferior vena cava (IVC).
To compare defibrillation thresholds (DFTs) of the PICD with those of a conventional ICD in humans.
Ten patients with ischemic cardiomyopathy and ejection fraction ≤35% were randomized to initial testing with either PICD or conventional ICD. A standard dual-coil lead was positioned in the RV apex. If randomized to PICD, the device was placed into the vasculature such that 1 titanium electrode was positioned in the SVC-BCV region and the second in the IVC. For PICD DFTs, the RV coil of the conventional ICD lead was connected to the PICD mandrel [shock vector: RV (+) to SVC-BCV (-) + IVC (-)]. When testing the conventional ICD, a subcutaneous pocket was formed in the left pectoralis region and the ICD was connected to the lead system and positioned in the pocket [shock vector: RV (+) to SVC (-) + active can (-)]. Each device was removed before testing with the other. A step-down binary search protocol determined the DFT, with the initial shock being 9 J.
The mean PICD DFT was 7.6 ± 3.3 J, and the conventional ICD system demonstrated a mean DFT of 9.5 ± 4.7 J (N = 10; paired t test, P = .28).
The intravascular defibrillator has DFTs similar to those of commercially available ICDs.
已研发出一种带有右心室(RV)单线圈导联和上腔静脉(SVC)-头臂静脉(BCV)区域及下腔静脉(IVC)内钛电极的经皮血管内除颤器(PICD)。
比较经皮血管内除颤器(PICD)与传统 ICD 在人体中的除颤阈值(DFT)。
10 例缺血性心肌病且射血分数≤35%的患者随机分为初始 PICD 或传统 ICD 测试组。将标准双线圈导联置于 RV 心尖部。若随机分至 PICD 组,则将设备置于血管中,使 1 个钛电极位于 SVC-BCV 区域,另 1 个位于 IVC。对于 PICD 的 DFT,将传统 ICD 导联的 RV 线圈连接至 PICD 轴[电击向量:RV(+)至 SVC-BCV(-)+IVC(-)]。当测试传统 ICD 时,在左侧胸肌区域形成皮下袋,并将 ICD 连接至导联系统并置于袋内[电击向量:RV(+)至 SVC(-)+主动罐(-)]。每次测试前均先移除另一个设备。采用逐步二进制搜索方案确定 DFT,初始电击强度为 9 J。