Muhney Kelly A, Dechow Paul C
CommuniCare Health Centers, Kyle, Texas, USA.
J Dent Hyg. 2010 Fall;84(4):185-9. Epub 2010 Nov 1.
To compare patients' perception of discomfort, vibration and noise levels between piezoelectric and the magnetostrictive ultrasonic units during periodontal debridement.
Periodontal debridement was performed on 75 subjects using a split-mouth design. Two quadrants on the same side were instrumented with a piezoelectric ultrasonic device (EMS Swiss Mini Master® Piezon) and the remaining 2 quadrants were instrumented with a magnetostrictive ultrasonic device (Dentsply Cavitron® SPS™). Subjects marked between 0 and 100 along a visual analog scale (VAS) for each of the 3 variables immediately after treatment of each half of the dentition. Scores of the VAS were compared using a nonparametric test for paired data, the Wilcoxon Signed-Rank test. The level of significance was set at p<0.05. Descriptive statistics included the median and the first and third quartiles as a measure of variation.
Mean scores for patient discomfort and vibration were greater for the magnetostrictive device at p=0.007 and p=0.032, respectively. The scores for noise level between the 2 ultrasonic types were almost equal.
The results show that, on average, patients in this study prefer instrumentation with the piezoelectric as it relates to awareness of associated discomfort and vibration. The results of this study may assist the clinician in the decision over which ultrasonic device may prove more beneficial in decreasing patient discomfort and increasing patient compliance.
比较在牙周清创过程中,患者对压电式和磁致伸缩式超声设备引起的不适、振动及噪音水平的感知。
采用双侧口腔设计,对75名受试者进行牙周清创。在同一侧的两个象限使用压电式超声设备(EMS Swiss Mini Master® Piezon)进行治疗,其余两个象限使用磁致伸缩式超声设备(Dentsply Cavitron® SPS™)进行治疗。在每侧牙列治疗后,受试者立即根据视觉模拟量表(VAS)对三个变量分别在0至100之间进行评分。使用配对数据的非参数检验(Wilcoxon符号秩检验)比较VAS评分。显著性水平设定为p<0.05。描述性统计包括中位数以及作为变异度量的第一和第三四分位数。
磁致伸缩式设备的患者不适和振动平均得分分别在p=0.007和p=0.032时更高。两种超声设备的噪音水平得分几乎相等。
结果表明,在本研究中,平均而言,患者更喜欢使用压电式设备进行治疗,因为其与相关不适和振动的感知有关。本研究结果可能有助于临床医生决定哪种超声设备在减轻患者不适和提高患者依从性方面可能更有益。