Invest Ophthalmol Vis Sci. 2011 Mar 10;52(3):1314-9. doi: 10.1167/iovs.10-5427.
To validate a new computer-aided diagnosis (CAD) imaging program for the assessment of nuclear lens opacity.
Slit-lamp lens photographs from the Singapore Malay Eye Study (SiMES) were graded using both the CAD imaging program and manual assessment method by a trained grader using the Wisconsin Cataract Grading System. Cataract was separately assessed clinically during the study using Lens Opacities Classification System III (LOCS III). The repeatability of CAD and Wisconsin grading methods were assessed using 160 paired images. The agreement between the CAD and Wisconsin grading methods, and the correlations of CAD with Wisconsin and LOCS III were assessed using the SiMES sample (5547 eyes from 2951 subjects).
In assessing the repeatability, the coefficient of variation (CoV) was 8.10% (95% confidence interval [CI], 7.21-8.99), and the intraclass correlation coefficient (ICC) was 0.96 (95% CI, 0.93-0.96) for the CAD method. There was high agreement between the CAD and Wisconsin methods, with a mean difference (CAD minus Wisconsin) of -0.02 (95% limit of agreement, -0.91 and 0.87) and an ICC of 0.81 (95% CI, 0.80-0.82). CAD parameters were also significantly correlated with LOCS III grading (all P < 0.001).
This new CAD imaging program assesses nuclear lens opacity with results comparable to the manual grading using the Wisconsin System. This study shows that an automated, precise, and quantitative assessment of nuclear cataract is possible.
验证一种新的计算机辅助诊断(CAD)成像程序,用于评估核性晶状体混浊。
使用新加坡马来眼研究(SiMES)的裂隙灯晶状体照片,通过训练有素的分级员使用威斯康星白内障分级系统(Wisconsin Cataract Grading System),同时使用 CAD 成像程序和手动评估方法进行分级。在研究期间,使用晶状体混浊分类系统 III(LOCS III)对白内障进行临床评估。使用 160 对图像评估 CAD 和威斯康星分级方法的重复性。使用 SiMES 样本(2951 名受试者的 5547 只眼)评估 CAD 与威斯康星分级方法的一致性,以及 CAD 与威斯康星和 LOCS III 的相关性。
在评估重复性时,CAD 方法的变异系数(CoV)为 8.10%(95%置信区间[CI],7.21-8.99),组内相关系数(ICC)为 0.96(95%CI,0.93-0.96)。CAD 方法与威斯康星方法具有高度一致性,平均差值(CAD 减去威斯康星)为-0.02(95%一致性界限,-0.91 和 0.87),ICC 为 0.81(95%CI,0.80-0.82)。CAD 参数与 LOCS III 分级也显著相关(均 P < 0.001)。
这种新的 CAD 成像程序可对核性晶状体混浊进行评估,其结果与使用 Wisconsin 系统的手动分级相当。本研究表明,对核性白内障进行自动、精确和定量评估是可行的。