Inserm U583, Institute for Neurosciences of Montpellier, CHU Saint-Eloi, Montpellier, France.
Ophthalmic Res. 2011;45(4):174-9. doi: 10.1159/000321118. Epub 2010 Nov 4.
Electroretinography (ERG) is a widely used technique to test retinal function in humans and animals. Recordings are particularly dependent on the type of electrodes used, with the best electrodes often being expensive and not always easy to use. The need of a simple and effective electrode type has led us to search the efficacy of different types of electrodes used in practice and compare them with the modified cotton wick electrode.
A modified type of electrode made of a cotton wick and impregnated with NaCl is described, and the ERG results were compared with other types of electrodes.
Compared with standard metal wire loop electrodes, the cotton wick electrode results in obtaining higher amplitudes, a better inter-eye correlation in the same animal and a better reproducibility of the recordings over time.
This cotton electrode is simple to make and easy to place. It provides reliable recordings during the entire life span of the animal and reliable comparisons between contralateral eyes, thus providing a powerful tool for ERG studies.
视网膜电图(ERG)是一种广泛用于测试人类和动物视网膜功能的技术。记录特别依赖于所使用的电极类型,而最好的电极通常价格昂贵且并不总是易于使用。因此,我们需要寻找一种简单有效的电极类型,我们研究了在实践中使用的不同类型的电极,并将它们与改良的棉线电极进行了比较。
描述了一种由棉线制成并浸有 NaCl 的改良型电极,并且将 ERG 结果与其他类型的电极进行了比较。
与标准的金属丝环电极相比,棉线电极可获得更高的振幅,在同一动物中具有更好的双眼间相关性,并且随着时间的推移具有更好的记录重现性。
这种棉电极制作简单,易于放置。它在动物的整个生命周期内提供可靠的记录,并在对侧眼睛之间提供可靠的比较,因此为 ERG 研究提供了有力的工具。