Rosner M, Bobak P, Lam T L
Georgiana Dvorak Theobald Ophthalmic Pathology Laboratory, Department of Ophthalmology and Visual Sciences, University of Illinois, Chicago College of Medicine.
Doc Ophthalmol. 1993;83(2):175-80. doi: 10.1007/BF01206215.
The corneal contact lens electrode, because of its convenience, has replaced the cotton-wick electrode for recording electroretinograms from patients and animals such as dogs, rabbits, and cats. The cotton-wick electrode, however, remains popular for rat electroretinogram measurements because small contact lens corneal electrodes that fit rat eyes are difficult to fabricate. We prepared corneal electrodes from disposable needles for use in recording electroretinograms from rats. The electrodes were readily prepared, were inexpensive, and were used successfully in six rats.
角膜接触镜电极因其便利性,已取代棉芯电极用于记录患者以及狗、兔、猫等动物的视网膜电图。然而,棉芯电极在大鼠视网膜电图测量中仍然很受欢迎,因为适合大鼠眼睛的小型角膜接触镜电极很难制作。我们用一次性针头制备了角膜电极,用于记录大鼠的视网膜电图。这些电极制备简便、价格低廉,并成功应用于六只大鼠。