Smith S J, Henderson L O, Hannon W H, Cooper G R
Division of Environmental Health Laboratory Sciences, Centers for Disease Control, Atlanta, GA 30333.
Clin Chem. 1990 Feb;36(2):290-6.
In 1987 a collaborative study was initiated with 140 laboratories worldwide to evaluate the effects of analytical method and lyophilization on the measurement of different concentrations of apolipoproteins (apo) A-I and B in four lyophilized serum pool samples. This survey confirmed that the lyophilized apo Reference Material of the International Union of Immunological Societies (IUIS) is useful for apo A-I assays as an international serum-based reference material, because among-method variation is negligible. The apo A-I concentration value of 1.24 g/L is now assigned to the IUIS Reference Material (CDC 1883) by a Centers for Disease Control RIA in-house reference method. Use of lyophilized serum preparation as a reference material for some modes of apo B measurement is questionable because of lyophilization and matrix effects. Both radial immunodiffusion and liquid immunoprecipitin methods demonstrated bias in measured apo B concentrations, compared with overall method-weighted means values on the IUIS Reference Material. Because of the uncertainty associated with LDL primary standard, protein analysis, and concentration differences among analytical methods, assigning a single apo B concentration value to the IUIS Reference Material appears inadvisable at present.
1987年,一项与全球140个实验室合作开展的研究启动,旨在评估分析方法和冻干处理对4份冻干血清混合样本中不同浓度载脂蛋白(apo)A-I和B测量结果的影响。该调查证实,国际免疫学会联盟(IUIS)的冻干载脂蛋白参考物质作为一种基于血清的国际参考物质,对于apo A-I检测是有用的,因为方法间差异可忽略不计。目前,疾病控制中心的放射免疫分析内部参考方法将1.24 g/L的apo A-I浓度值赋予了IUIS参考物质(CDC 1883)。由于冻干和基质效应,将冻干血清制剂用作某些apo B测量模式的参考物质存在疑问。与IUIS参考物质上的总体方法加权均值相比,放射免疫扩散法和液体免疫沉淀法在测量apo B浓度时均显示出偏差。由于与低密度脂蛋白一级标准、蛋白质分析以及分析方法之间的浓度差异相关的不确定性,目前给IUIS参考物质指定单一的apo B浓度值似乎并不可取。