Suppr超能文献

利用光谱衰减全反射椭圆偏振技术测量红外波段液体的复折射率:散射去偏振校正

Measurement of the complex refractive index of liquids in the infrared using spectroscopic attenuated total reflection ellipsometry: correction for depolarization by scattering.

作者信息

den Boer J H, Kroesen G M, de Hoog F J

出版信息

Appl Opt. 1995 Sep 1;34(25):5708-14. doi: 10.1364/AO.34.005708.

Abstract

With spectroscopic ellipsometry one can measure the real and imaginary parts of the refractive index of a medium simultaneously. To determine this index in the infrared for a number of technical liquids, use was made of attenuated total internal reflection at the glass-liquid interface of a specially designed prism. This attenuated total reflection approach warrants minimal signal loss and is, for strongly absorbing liquids, the only way to measure the complex refractive index. A surprising phenomenon, observed when BK-7 prism glass was used, is scattering in the vicinity of the absorption wavelengths of the glass. A simple model that can be used to describe the relations among absorption, scattering, and depolarization was successfully used to correct the measurements. Refractive indices for demineralized water, Freon 113, heptane, benzene, gas oil, and crude oil in the wave number range from 5000 to 10,000 cm(-1) (1-2 µm) are presented.

摘要

利用椭圆偏振光谱法可以同时测量介质折射率的实部和虚部。为了测定多种工业液体在红外波段的折射率,采用了在特殊设计的棱镜的玻璃 - 液体界面处的衰减全内反射。这种衰减全反射方法保证了最小的信号损失,对于强吸收性液体而言,是测量复折射率的唯一方法。当使用BK - 7棱镜玻璃时,观察到一个令人惊讶的现象,即在玻璃吸收波长附近出现散射。一个可用于描述吸收、散射和去极化之间关系的简单模型被成功用于校正测量结果。给出了去离子水、氟利昂113、庚烷、苯、粗柴油和原油在波数范围从5000至10,000 cm⁻¹(1 - 2 µm)的折射率。