Sansonetti C J, Salit M L, Reader J
Appl Opt. 1996 Jan 1;35(1):74-7. doi: 10.1364/AO.35.000074.
The wavelengths of 19 spectral lines in the region 253-579 nm emitted by Hg pencil-type lamps were measured by Fourier-transform spectroscopy. Precise calibration of the spectra was obtained with wavelengths of (198)Hg as external standards. Our recommended values should be useful aswavelength-calibration standards for moderate-resolution spectrometers at an uncertainty level of 0.0001 nm.
利用傅里叶变换光谱法测量了汞笔型灯发射的253 - 579 nm区域内19条光谱线的波长。以(198)汞的波长作为外标对光谱进行了精确校准。我们推荐的值作为中等分辨率光谱仪的波长校准标准,不确定度水平为0.0001 nm,应该会很有用。