与传统鼻持续气道正压通气相比,新生儿头盔式持续气道正压通气时的噪音暴露增加。

Noise exposure is increased with neonatal helmet CPAP in comparison with conventional nasal CPAP.

作者信息

Trevisanuto D, Camiletti L, Doglioni N, Cavallin F, Udilano A, Zanardo V

机构信息

Departments of Pediatric, Medical School, University of Padua, Azienda Ospedaliera di Padova, Padua, Italy.

出版信息

Acta Anaesthesiol Scand. 2011 Jan;55(1):35-8. doi: 10.1111/j.1399-6576.2010.02356.x. Epub 2010 Nov 15.

Abstract

BACKGROUND

in adults, noninvasive ventilation via a helmet is associated with significantly greater noise than nasal and facial masks. We hypothesized that noise exposure could be increased with neonatal helmet continuous positive airway pressure (CPAP) in comparison with conventional nasal CPAP (nCPAP). Our primary objective was to compare the noise intensity produced by a neonatal helmet CPAP and a conventional nCPAP system. Furthermore, we aimed to evaluate the effect of the gas flow rate and the presence of the humidifier and the filter on noise levels during neonatal helmet CPAP treatment.

METHODS

in this bench study, noise intensity was measured in the following settings: helmet CPAP, nCPAP, incubator and the neonatal intensive care unit. In helmet CPAP, noise measurements were performed at different gas flow rates (8, 10 and 12 l/min), while in nCPAP, the flow rate was 8 l/min. For both CPAP systems, the level of pressure was maintained constant at 5 cmH(2) O.

RESULTS

during neonatal helmet CPAP, the median (interquartile range) noise levels were significantly higher than those during nCPAP: 70.0 dB (69.9-70.4) vs. 62.7 dB (62.5-63.0); P<0.001. In the helmet CPAP, the noise intensities changed with increasing flow rate and with the presence of a humidifier or a filter.

CONCLUSIONS

noise intensities generated by the neonatal helmet CPAP were significantly higher than those registered while using a conventional nCPAP system. In the helmet, the noise intensity depends on the gas flow rate, and the presence of a humidifier and a filter in the system.

摘要

背景

在成人中,通过头盔进行无创通气所产生的噪音明显高于鼻罩和面罩。我们推测,与传统鼻持续气道正压通气(nCPAP)相比,新生儿头盔持续气道正压通气(CPAP)可能会增加噪音暴露。我们的主要目标是比较新生儿头盔CPAP和传统nCPAP系统产生的噪音强度。此外,我们旨在评估新生儿头盔CPAP治疗期间气体流速以及加湿器和过滤器的存在对噪音水平的影响。

方法

在这项实验台研究中,在以下环境中测量噪音强度:头盔CPAP、nCPAP、 incubator和新生儿重症监护病房。在头盔CPAP中,在不同气体流速(8、10和12升/分钟)下进行噪音测量,而在nCPAP中,流速为8升/分钟。对于两种CPAP系统,压力水平均保持在5 cmH₂O不变。

结果

在新生儿头盔CPAP期间,中位数(四分位间距)噪音水平显著高于nCPAP期间:70.0分贝(69.9 - 70.4)对62.7分贝(62.5 - 63.0);P < 0.001。在头盔CPAP中,噪音强度随流速增加以及加湿器或过滤器的存在而变化。

结论

新生儿头盔CPAP产生的噪音强度显著高于使用传统nCPAP系统时记录的强度。在头盔中,噪音强度取决于气体流速以及系统中加湿器和过滤器的存在。