Jagomägi Kersti, Talts Jaak, Tähepõld Peeter, Raamat Rein
Department of Physiology, University of Tartu, Estonia.
Clin Physiol Funct Imaging. 2011 May;31(3):188-92. doi: 10.1111/j.1475-097X.2010.00997.x. Epub 2010 Nov 15.
Non-invasive beat-to-beat mean arterial pressure (MAP) in finger arteries recorded by the differential oscillometric device was compared with MAP recorded invasively from A. radialis in 22 patients after cardiac surgery. Based on all 132 paired measurements, the MAP values measured at the radial artery were 2.7 ± 4.9 mmHg higher than those measured on fingers. Among 22 patients there were 8 patients receiving inotropic support, their difference being 2.1 ± 5.6 mmHg. The present study revealed that the mean discrepancy between the invasive radial pressure and finger pressure was small; however, patient data sets showed marked variability in average pressure differences when examined individually.
通过示波法测量装置记录的手指动脉无创逐搏平均动脉压(MAP)与22例心脏手术后患者桡动脉有创记录的MAP进行比较。基于全部132对测量值,桡动脉测量的MAP值比手指测量的MAP值高2.7±4.9mmHg。22例患者中有8例接受了正性肌力支持,其差值为2.1±5.6mmHg。本研究表明,有创桡动脉压与手指压之间的平均差异较小;然而,对患者数据集进行单独检查时,平均压差存在显著变异性。