Radiation Oncology, Royal Prince Alfred Hospital, New South Wales, Australia.
Int J Radiat Oncol Biol Phys. 2011 Feb 1;79(2):609-15. doi: 10.1016/j.ijrobp.2010.03.030. Epub 2010 Nov 18.
To report on the clinical feasibility of a novel scintillation detector system with fiberoptic readout that measures the urethral dose during high-dose-rate brachytherapy treatment of the prostate.
The clinical trial enrolled 24 patients receiving high-dose-rate brachytherapy treatment to the prostate. After the first 14 patients, three improvements were made to the dosimeter system design to improve clinical reliability: a dosimeter self-checking facility; a radiopaque marker to determine the position of the dosimeter, and a more robust optical extension fiber.
Improvements to the system design allowed for accurate dose measurements to be made in vivo. A maximum measured dose departure of 9% from the calculated dose was observed after dosimeter design improvements.
Departures of the measured from the calculated dose, after improvements to the dosimetry system, arise primarily from small changes in patient anatomy. Therefore, we recommend that patient response be correlated with the measured in vivo dose rather than with the calculated dose.
报告一种新型闪烁探测器系统的临床可行性,该系统采用光纤读出技术,可测量前列腺高剂量率近距离治疗过程中的尿道剂量。
该临床试验纳入了 24 例接受前列腺高剂量率近距离治疗的患者。在前 14 例患者后,对剂量计系统设计进行了三项改进,以提高临床可靠性:剂量计自检功能;不透射线标记物,以确定剂量计的位置;以及更坚固的光学延伸光纤。
系统设计的改进使得可以进行体内准确的剂量测量。在剂量计设计改进后,观察到最大实测剂量偏离计算剂量 9%。
在对剂量计系统进行改进后,实测剂量与计算剂量的偏差主要源于患者解剖结构的微小变化。因此,我们建议将患者的反应与体内实测剂量相关联,而不是与计算剂量相关联。