PKPD Core Facility, School of Pharmacy, Queens University Belfast, 97 Lisburn Road, Belfast, United Kingdom.
J Pharm Biomed Anal. 2011 Mar 25;54(4):701-10. doi: 10.1016/j.jpba.2010.10.025. Epub 2010 Nov 3.
A new stir bar sorptive extraction (SBSE) technique coupled with HPLC-UV method for quantification of diclofenac in pharmaceutical formulations has been developed and validated as a proof of concept study. Commercially available polydimethylsiloxane stir bars (Twister™) were used for method development and SBSE extraction (pH, phase ratio, stirring speed, temperature, ionic strength and time) and liquid desorption (solvents, desorption method, stirring time etc) procedures were optimised. The method was validated as per ICH guidelines and was successfully applied for the estimation of diclofenac from three liquid formulations viz. Voltarol(®) Optha single dose eye drops, Voltarol(®) Ophtha multidose eye drops and Voltarol(®) ampoules. The developed method was found to be linear (r=0.9999) over 100-2000ng/ml concentration range with acceptable accuracy and precision (tested over three QC concentrations). The SBSE extraction recovery of the diclofenac was found to be 70% and the LOD and LOQ of the validated method were found to be 16.06 and 48.68ng/ml, respectively. Furthermore, a forced degradation study of a diclofenac formulation leading to the formation of structurally similar cyclic impurity (indolinone) was carried out. The developed extraction method showed comparable results to that of the reference method, i.e. method was capable of selectively extracting the indolinone and diclofenac from the liquid matrix. Data on inter and intra stir bar accuracy and precision further confirmed robustness of the method, supporting the multiple re-use of the stir bars.
一种新的搅拌棒吸附萃取 (SBSE) 技术与 HPLC-UV 方法相结合,已被开发并验证为概念验证研究,用于定量测定药物制剂中的双氯芬酸。商业上可获得的聚二甲基硅氧烷搅拌棒 (Twister™) 用于方法开发和 SBSE 萃取 (pH 值、相比、搅拌速度、温度、离子强度和时间) 和液相解吸 (溶剂、解吸方法、搅拌时间等) 程序进行优化。该方法按照 ICH 指南进行验证,并成功应用于三种液体制剂中双氯芬酸的估计,即 Voltrol(®) Optha 单剂量眼药水、Voltrol(®) Ophtha 多剂量眼药水和 Voltrol(®) 安瓿。所开发的方法在 100-2000ng/ml 浓度范围内具有良好的线性关系(r=0.9999),并具有可接受的准确性和精密度(在三个 QC 浓度下进行测试)。双氯芬酸的 SBSE 萃取回收率为 70%,验证方法的 LOD 和 LOQ 分别为 16.06 和 48.68ng/ml。此外,还对一种导致结构相似的环状杂质(吲哚啉酮)形成的双氯芬酸制剂进行了强制降解研究。所开发的提取方法与参考方法相比,表现出相当的结果,即该方法能够从液体基质中选择性地提取吲哚啉酮和双氯芬酸。关于搅拌棒之间和内部准确性和精密度的数据进一步证实了方法的稳健性,支持搅拌棒的多次重复使用。