Suppr超能文献

基于肌电图模式识别的假肢控制中采样率的选择。

Selection of sampling rate for EMG pattern recognition based prosthesis control.

作者信息

Li Guanglin, Li Yaonan, Zhang Zhiyong, Geng Yanjuan, Zhou Rui

机构信息

Institute of Biomedical and Health Engineering, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Guangdong 518055 China.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:5058-61. doi: 10.1109/IEMBS.2010.5626224.

Abstract

Most previous studies of electromyography (EMG) pattern recognition control of multifunctional myoelectric prostheses adopted a conventional sampling rate that is commonly used in EMG research fields. However, it is unknown whether using a lower sampling rate in EMG acquisition still preserves sufficient neural control information for accurate classification of user movement intents. This study investigated the effects of EMG sampling rate on the performance of EMG pattern recognition in identifying 11 classes of arm and hand movements. Our results showed that decreasing the sampling rate from 1 kHz to 500 Hz only caused 0.8% reduction of the average classification accuracy over five able-bodied subjects and 2.2% decrease over two transradial amputees. When using a 400 Hz sampling rate, the average classification accuracy decreased 1.3% and 2.8% in able-bodied subjects and amputees, respectively. These results suggest that a sampling rate between 400-500 Hz would be optimal for EMG acquisition in EMG pattern recognition based control of a multifunctional prosthesis.

摘要

此前大多数关于多功能肌电假肢肌电图(EMG)模式识别控制的研究都采用了EMG研究领域常用的传统采样率。然而,在EMG采集过程中使用较低的采样率是否仍能保留足够的神经控制信息以准确分类用户的运动意图,目前尚不清楚。本研究调查了EMG采样率对识别11种手臂和手部运动的EMG模式识别性能的影响。我们的结果表明,将采样率从1 kHz降低到500 Hz,在五名健全受试者中平均分类准确率仅降低了0.8%,在两名经桡骨截肢者中降低了2.2%。当使用400 Hz采样率时,健全受试者和截肢者的平均分类准确率分别下降了1.3%和2.8%。这些结果表明,在基于EMG模式识别的多功能假肢控制中,400 - 500 Hz之间的采样率对于EMG采集而言是最佳的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验