Takahashi Masa, Kinase Sakae, Kramer Richard
Japan Atomic Energy Agency, 2-4 Shirakata-Shirane, Tokai-mura, Ibaraki 319-1195, Japan.
Radiat Prot Dosimetry. 2011 Mar;144(1-4):407-10. doi: 10.1093/rpd/ncq417. Epub 2010 Dec 4.
In the 2007 recommendations of the International Commission on Radiological Protection, it is mentioned that the reference voxel phantoms are used for calculation of effective dose. From the standpoint of internal dosimetry services, calibration methods of whole-body counters using the voxel phantoms are of considerable practical interest. In the present study, counting efficiencies of a whole-body counter at the Japan Atomic Energy Agency (JAEA) were evaluated using Monte Carlo simulations with two voxel phantoms, 'MAX06', which has organ masses corresponding to those of the reference male, and 'Otoko', which is a representation for average Japanese male. To validate the calculation methods of the present study, calculations for the bottle manikin absorption phantom were also performed and compared with experiments. Consequently, it was found that the Monte Carlo simulation with voxel phantoms is a significant tool for the calibration of the JAEA whole-body counter.
在国际放射防护委员会2007年的建议中提到,参考体素模型用于有效剂量的计算。从内照射剂量测定服务的角度来看,使用体素模型的全身计数器校准方法具有相当大的实际意义。在本研究中,利用蒙特卡罗模拟,使用两个体素模型对日本原子能机构(JAEA)的全身计数器的计数效率进行了评估,这两个模型分别是“MAX06”(其器官质量与参考男性的器官质量相对应)和“Otoko”(代表日本男性平均水平)。为了验证本研究的计算方法,还对瓶状人体模型吸收体模进行了计算并与实验进行了比较。结果发现,使用体素模型的蒙特卡罗模拟是校准JAEA全身计数器的重要工具。