Kinase S, Takagi S, Noguchi H, Saito K
Japan Atomic Energy Agency 2-4 Shirane, Shirakata, Tokai-mura, Ibaraki 319-1195, Japan.
Radiat Prot Dosimetry. 2007;125(1-4):189-93. doi: 10.1093/rpd/ncm197. Epub 2007 May 22.
Voxel phantoms and Monte Carlo methods are of considerable interest in calibrating in vivo counting applications for radioactive body burden such as whole-body counters. At the Japan Atomic Energy Research Institute (JAERI), a calculation code--UCWBC code--for whole-body counter calibrations using voxel phantoms has been developed as an EGS4 Monte Carlo user code. To validate the UCWBC code for calibrating whole-body counters at JAERI, response functions and counting efficiencies of a p-type high-purity Ge (HPGe) semiconductor detector used for the whole-body counter were evaluated for a water-filled block-shape phantom by use of UCWBC code and were measured by experiments. The voxel version of the water-filled block-shape phantom based on an actual phantom was used for the calculations. Furthermore, counting efficiencies of the Ge semiconductor detector for the male and female voxel phantoms developed at JAERI were evaluated in the photon energy range from 60 to 1836 keV by the UCWBC code in order to examine the differences between the counting efficiencies for voxel phantoms. In conclusion, it was found that the response functions and counting efficiencies of the Ge semiconductor detector found by the UCWBC code for the water-filled block-shape phantom are in good agreement with measured data. The UCWBC code was validated by those comparisons. It was also found that the counting efficiencies of the Ge semiconductor detector depend on the size of the phantoms and the effective distance between phantom and detector. The calibration of whole-body counters using voxel phantoms and Monte Carlo methods would be quite useful for the improvement in the accuracy of measurement results.
体素模型和蒙特卡罗方法在校准用于测量放射性体内负荷(如全身计数器)的体内计数应用方面具有相当大的意义。在日本原子能研究所(JAERI),已开发出一种用于使用体素模型进行全身计数器校准的计算代码——UCWBC代码,作为EGS4蒙特卡罗用户代码。为了验证JAERI的UCWBC代码用于校准全身计数器的有效性,使用UCWBC代码对用于全身计数器的p型高纯锗(HPGe)半导体探测器的响应函数和计数效率进行了评估,并通过实验进行了测量。基于实际模型的充水块状体素模型用于计算。此外,为了研究体素模型计数效率之间的差异,通过UCWBC代码在60至1836 keV的光子能量范围内评估了JAERI开发的男性和女性体素模型的锗半导体探测器的计数效率。总之,发现UCWBC代码得出的充水块状体素模型的锗半导体探测器的响应函数和计数效率与测量数据高度吻合。通过这些比较验证了UCWBC代码。还发现锗半导体探测器的计数效率取决于模型的大小以及模型与探测器之间的有效距离。使用体素模型和蒙特卡罗方法对全身计数器进行校准对于提高测量结果的准确性将非常有用。