Rapid Commun Mass Spectrom. 2011 Jan 15;25(1):140-6. doi: 10.1002/rcm.4836.
A rotating ball interface for surface-assisted laser desorption/ionization (SALDI) mass spectrometry was designed and tested. One side of the ball was exposed to atmospheric pressure and the other to the vacuum in a time-of-flight mass spectrometer. Analytes (arginine, atenolol, reserpine, tofisopam, and chloropyramine) were applied using electrospray to a silicon substrate on the atmospheric side, the ball was rotated 180°, and the analyte was desorbed on the vacuum side using a pulsed, 200 Hz, 355 nm laser. In order to increase the desorption area, the laser focus was scanned over the substrate in a raster pattern repeated once every second. The design allows for rapid sample throughout with a sample turn-around time as short as 5 s. Newly produced porous silicon substrates initially yielded very low ion signals, and they required several hundred laser shots to attain maximum sensitivity. In contrast, amorphous silicon did not require such 'activation'. Quantitative analysis showed a sample-to-sample reproducibility of about 10%. The sensitivities with model analytes were in the 1000 to 10,000 ions/fmole range and detection limits in the low fg range.
设计并测试了用于表面辅助激光解吸/电离 (SALDI) 质谱的旋转球接口。球的一侧暴露在大气压下,另一侧暴露在飞行时间质谱仪的真空中。分析物(精氨酸、阿替洛尔、利血平、甲苯噻嗪和氯吡拉明)通过电喷雾施加到大气压侧的硅衬底上,将球旋转 180°,并用脉冲、200 Hz、355nm 激光从真空侧解吸分析物。为了增加解吸面积,激光焦点以每秒重复一次的光栅图案在衬底上扫描。该设计允许快速进行样品处理,样品周转时间最短可达 5 秒。新生产的多孔硅衬底最初产生的离子信号非常低,需要数百次激光照射才能达到最大灵敏度。相比之下,非晶硅不需要这种“激活”。定量分析显示,样品之间的重现性约为 10%。模型分析物的灵敏度在 1000 到 10000 离子/fmole 范围内,检测限在 fg 低范围内。