Gambini D-J
Service Central de Médecine du Travail, Hôtel Dieu, 1 place du Parvis Notre Dame, 75004 Paris, France.
J Radiol. 2010 Nov;91(11 Pt 2):1186-8. doi: 10.1016/s0221-0363(10)70171-2.
An x-ray tube mainly emits low-energy X-rays, with few maximum energy E₀ (equal in keV to the voltage U in kV) x-rays. Aluminium filtration (mandatory minimum thickness of 1.5 to 2.5 mm based on tube voltage) reduces soft X-rays and provides a mean energy equal to 2/3 E₀. The half value layer of a reference material characterizes the spectrum. X-ray attenuation in tissues is due to secondary electron interactions: photoelectric effect at low-energy, especially in dense materials with high Z number; compton effect at intermediate-energy, proportional to density. The optimization of acquisition parameters of a medically necessary examination is based on appropriate selection of the highest voltage (U in kV) providing the best contrast and lowest tube current (Q in mAs) providing a diagnostic image.