Comparison of the accuracy and precision of prostate localization with 2D-2D and 3D images.

机构信息

Department of Radiation Oncology, Rigshospitalet, Copenhagen University Hospital, Denmark.

出版信息

Radiother Oncol. 2011 Feb;98(2):175-80. doi: 10.1016/j.radonc.2010.11.012. Epub 2010 Dec 21.

Abstract

BACKGROUND AND PURPOSE

Positional uncertainties related to the set-up of the prostate, using internal markers and either 2D-2D or 3D images, were studied. Set-up using direct prostate localization on CBCT scans is compared to set-up using internal markers.

MATERIAL AND METHODS

20 patients with prostate cancer were enrolled in the study. After each daily session, a set of 2D-2D and 3D images were acquired. The images isocenter was compared to reference images isocenter. For the set-up error analysis the systematic error, μ, and the set-up uncertainties, Σ and σ, were determined for the translational shift in the three directions, lat, lng and vrt. The set-up errors and uncertainties were calculated in the same way for rotations around the three axes, lat, lng and vrt.

RESULTS

Set-up uncertainties were evaluated for four different set-up methods. The systematic error uncertainties were found to be in the range 0.38-1.14 mm and for the random error 0.79-1.48 mm. For rotations the uncertainties ranges were 0.38-1.59° and 0.91-2.18° for systematic and random uncertainties, respectively. Set-up uncertainties, using internal markers or prostate itself to position the target in the isocenter, were comparable. The correlation between the two methods was better for translational shifts of the isocenter than for rotational shifts.

CONCLUSIONS

The study shows that the precision of the 2D-2D set-up is equivalent to the precision of the 3D images. It also shows that the soft-tissue based set-up needs 1 mm larger set-up margins than the marker based set-up for the prostate patients, when CBCT is used for daily verification of the location of the prostate.

摘要

背景与目的

研究了使用内部标记物和二维-二维或三维图像进行前列腺设置时的位置不确定性。比较了直接在 CBCT 扫描上进行前列腺定位的设置和使用内部标记物的设置。

材料与方法

20 例前列腺癌患者纳入本研究。每天治疗后,采集一组二维-二维和三维图像。将图像等中心点与参考图像等中心点进行比较。对于定位误差分析,确定了三个方向(lat、lng 和 vrt)的平移位移的系统误差μ和定位不确定性Σ和σ。以相同的方式计算了三个轴(lat、lng 和 vrt)的旋转的定位误差和不确定性。

结果

评估了四种不同定位方法的定位不确定性。系统误差不确定性在 0.38-1.14mm 范围内,随机误差在 0.79-1.48mm 范围内。旋转时,系统不确定性范围为 0.38-1.59°,随机不确定性范围为 0.91-2.18°。使用内部标记物或前列腺本身来将目标定位在等中心时,定位不确定性相当。对于等中心的平移位移,这两种方法之间的相关性优于旋转位移。

结论

本研究表明,二维-二维定位的精度与三维图像的精度相当。它还表明,当使用 CBCT 对前列腺位置进行日常验证时,与基于标记物的设置相比,基于软组织的设置需要 1mm 更大的设置余量。