Boziari A, Koukorava C, Carinou E, Hourdakis C J, Kamenopoulou V
Greek Atomic Energy Commission, P Grigoriou & Neapoleos, Ag Paraskevi, 15310 Athens, Greece.
Radiat Prot Dosimetry. 2011 Mar;144(1-4):173-6. doi: 10.1093/rpd/ncq545. Epub 2010 Dec 31.
The use of active personal dosemeters (APDs) not only as a warning device but also, in some cases, as an official and hence stand-alone dosemeter is rapidly increasing. A comparison in terms of dose, energy and angle dependence, among different types of APD and a routinely used whole-body thermoluminescence dosemeter (TLD) has been performed. Significant differences were found between the TLD readings and mainly some not commonly used APDs. The importance of choosing the best adapted APD according to the radiation field characteristics is pointed out.
主动式个人剂量计(APD)不仅用作警示设备,而且在某些情况下还作为官方的独立剂量计使用,其应用正在迅速增加。我们对不同类型的APD与常规使用的全身热释光剂量计(TLD)在剂量、能量和角度依赖性方面进行了比较。结果发现,TLD读数与主要一些不常用的APD之间存在显著差异。文中指出了根据辐射场特性选择最适配APD的重要性。