Development of patient support devices for execution of clinical radiotherapy for cancer patients: A preliminary report.

作者信息

Babu N K, Singh Bakshish, Namrata S, Mohanti B K, Ravichandran R, Ghamrawy K E

机构信息

Department of Radiotherapy, National Oncology Center, Royal Hospital, Muscat, Sultanate of Oman.

出版信息

J Med Phys. 2006 Oct;31(4):255-61. doi: 10.4103/0971-6203.29195.

Abstract

The present paper illustrates our attempt to design and test the reproducibility of low-cost patient positioning devices prepared in-house in our radiotherapy department. Rigid thermocole boards with angulations, scales and support were designed as breast, pelvis and head positioning devices. Reproducibility and accuracy were tested by serial electronic portal imaging detector imaging.The positioning devices (with or without superimposed moulds) showed variations within 2-3 mm on serial treatment days which were within acceptable limits.It is therefore concluded that low-cost patient positioning devices for head, breast and pelvis (the common sites of treatments in radiotherapy) can be fabricated from available materials in-house. These have been shown to be resulting in accurate immobilization, can be customized for particular techniques and are considerably cheaper than commercially available solutions.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0ce0/3004100/5bad2458cd84/JMP-31-255-g001.jpg

相似文献

2
The use of a standardized positioning support cushion during daily routine of breast irradiation.
Int J Radiat Oncol Biol Phys. 1998 May 1;41(2):459-63. doi: 10.1016/s0360-3016(98)00022-4.
10
Immobilizing and Positioning Patients for Radiotherapy.放疗患者的固定与体位摆放
Semin Radiat Oncol. 1995 Apr;5(2):100-114. doi: 10.1054/SRAO00500100.

本文引用的文献

3
Comparison of two head and neck immobilization systems.两种头颈部固定系统的比较。
Int J Radiat Oncol Biol Phys. 1997 Jul 1;38(4):867-73. doi: 10.1016/s0360-3016(97)00075-8.
4
The effectiveness of immobilization during prostate irradiation.
Int J Radiat Oncol Biol Phys. 1995 Jan 1;31(1):143-8. doi: 10.1016/0360-3016(94)00351-K.
5
Automatic verification of radiation field shape using digital portal images.
Med Phys. 1992 Jul-Aug;19(4):1007-14. doi: 10.1118/1.596818.
7
Immobilization error: some theoretical considerations.
Radiology. 1975 Nov;117(2):407-12. doi: 10.1148/117.2.407.
8
The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields.
Cancer. 1976 Jun;37(6):2755-61. doi: 10.1002/1097-0142(197606)37:6<2755::aid-cncr2820370628>3.0.co;2-j.
9
Weekly localization films and detection of field placement errors.每周进行定位片检查并检测野放置误差。
Int J Radiat Oncol Biol Phys. 1978 Sep-Oct;4(9-10):881-7. doi: 10.1016/0360-3016(78)90051-2.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验