[Echodoppler evaluation of the normally functioning Saint-Jude's aortic valve prosthesis].

作者信息

Lesbre J P, Guillaumont M P, Dallocchio M, Roudaut R, Tribouilloy C, Choquet D

机构信息

Service de cardiologie B, CHU, Amiens.

出版信息

Arch Mal Coeur Vaiss. 1990 Sep;83(10):1553-61.

PMID:2122832
Abstract

The aims of this study were: to define Doppler echocardiographic criteria of normality of aortic St Jude Medical (SJM) valve prostheses with respect to their size and to verify the validity of the continuity equation in the determination of prosthetic valve functional surface area. Forty patients with apparently normally functioning SJM prostheses without other cardiac disease were investigated at least one month after surgery. The group consisted in 1 n. 19, 6 n. 21, 9 n. 23, 12 n. 25 and 12 n. 27 SJM prostheses. The following parameters were measured: the maximum transprosthetic velocity, maximum and mean transprosthetic pressure gradients, permeability index and the Doppler surface area calculated by the continuity equation using the method proposed by Skjaerpe. The global results were as follows: maximum velocity = 2.5 +/- 0.4 m/s (1.8-3.7 m/s); maximum gradient = 26.9 +/- 9.8 mmHg (14-53 mmHg); mean gradient = 13.7 +/- 5.6 mmHg (7-30 mmHg); permeability index = 0.41 +/- 0.09 (0.23-0.57); Doppler surface area = 1.89 +/- 0.66 cm2 (0.73-3.23 cm2). When the prostheses were considered according to their sizes a weak negative correlation was observed between the mean pressure gradients and the size of the prostheses: r = -0.43, p less than 0.05 and a positive correlation between Doppler surface area and the theoretical prosthetic surface area: r = 0.71, p less than 0.005; SD = 0.45 cm2. No significant differences were observed between the pressure gradients and velocities of each size of prosthesis except when sizes 21 + 23 were compared with the large sizes (n. 25 + 27).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

本研究的目的是

确定圣犹达医疗公司(SJM)主动脉瓣膜假体正常的多普勒超声心动图标准,涉及假体大小,并验证连续性方程在确定人工瓣膜功能表面积方面的有效性。对40例无其他心脏疾病且SJM假体功能明显正常的患者在术后至少1个月进行了研究。该组包括1例型号19、6例型号21、9例型号23、12例型号25和12例型号27的SJM假体。测量了以下参数:最大跨假体速度、最大和平均跨假体压力梯度、通透指数以及使用Skjaerpe提出的方法通过连续性方程计算的多普勒表面积。总体结果如下:最大速度 = 2.5±0.4 m/s(1.8 - 3.7 m/s);最大梯度 = 26.9±9.8 mmHg(14 - 53 mmHg);平均梯度 = 13.7±5.6 mmHg(7 - 30 mmHg);通透指数 = 0.41±0.09(0.23 - 0.57);多普勒表面积 = 1.89±0.66 cm²(0.73 - 3.23 cm²)。根据假体大小进行分析时,观察到平均压力梯度与假体大小之间存在微弱的负相关:r = -0.43,p < 0.05;多普勒表面积与理论假体表面积之间存在正相关:r = 0.71,p < 0.005;标准差 = 0.45 cm²。除了将型号21 + 23与大尺寸型号(型号25 + 27)进行比较时,各尺寸假体的压力梯度和速度之间未观察到显著差异。(摘要截断于250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验