Suppr超能文献

质谱分析系统的小型化

Miniaturization of Mass Spectrometry Analysis Systems.

作者信息

Xu Wei, Manicke Nicholas E, Cooks Graham R, Ouyang Zheng

机构信息

Weldon School of Biomedical Engineering, Purdue University, West Lafayette, IN 47907, USA.

出版信息

JALA Charlottesv Va. 2010 Dec;15(6):433-439. doi: 10.1016/j.jala.2010.06.004.

Abstract

The key concepts and technologies developed in our laboratories in Purdue University for the miniaturization of mass spectrometry analysis systems are introduced. Mass analyzers of simple geometries with a novel atmospheric pressure interface were employed allowed reduction in the size of the ion trap mass spectrometer. Ambient ionization methods were developed and coupled to miniature mass spectrometers to allow direct MS analysis of complex samples without sample preparation and chemical separation. The performance of desorption electrospray ionization, low temperature plasma probe, paper spray as well as two handheld MS systems, Mini 10 and Mini 11, are described with demonstrations of capabilities for chemical analysis.

摘要

本文介绍了我们在普渡大学实验室中为质谱分析系统小型化所开发的关键概念和技术。采用了具有新型大气压接口的简单几何形状的质量分析器,从而减小了离子阱质谱仪的尺寸。开发了常压电离方法并将其与微型质谱仪相结合,以实现对复杂样品的直接质谱分析,无需样品制备和化学分离。文中描述了解吸电喷雾电离、低温等离子体探针、纸喷雾以及两款手持式质谱系统Mini 10和Mini 11的性能,并展示了它们的化学分析能力。

相似文献

1
Miniaturization of Mass Spectrometry Analysis Systems.
JALA Charlottesv Va. 2010 Dec;15(6):433-439. doi: 10.1016/j.jala.2010.06.004.
2
Sampling wand for an ion trap mass spectrometer.
Anal Chem. 2011 Mar 1;83(5):1857-61. doi: 10.1021/ac102962e. Epub 2011 Feb 7.
3
[Detection of drugs in urine by ambient direct ionization mass spectrometry].
Se Pu. 2022 Jul;40(7):677-683. doi: 10.3724/SP.J.1123.2022.01013.
4
Breaking the pumping speed barrier in mass spectrometry: discontinuous atmospheric pressure interface.
Anal Chem. 2008 Jun 1;80(11):4026-32. doi: 10.1021/ac800014v. Epub 2008 May 8.
6
Utility of mass spectrometry for proteome analysis: part I. Conceptual and experimental approaches.
Expert Rev Proteomics. 2008 Dec;5(6):841-64. doi: 10.1586/14789450.5.6.841.
7
Design and characterization of a multisource hand-held tandem mass spectrometer.
Anal Chem. 2008 Oct 1;80(19):7198-205. doi: 10.1021/ac801275x. Epub 2008 Aug 29.
8
[Advances in thin layer chromatography coupled with mass spectrometry technology].
Se Pu. 2023 Jan;41(1):24-36. doi: 10.3724/SP.J.1123.2022.03038.
9
Surface analysis of lipids by mass spectrometry: more than just imaging.
Prog Lipid Res. 2013 Oct;52(4):329-53. doi: 10.1016/j.plipres.2013.04.005. Epub 2013 Apr 24.
10
Mini 12, miniature mass spectrometer for clinical and other applications--introduction and characterization.
Anal Chem. 2014 Mar 18;86(6):2909-16. doi: 10.1021/ac403766c. Epub 2014 Feb 24.

引用本文的文献

2
Faster, better, and cheaper: harnessing microfluidics and mass spectrometry for biotechnology.
RSC Chem Biol. 2021 Jul 20;2(5):1331-1351. doi: 10.1039/d1cb00112d. eCollection 2021 Oct 7.
4
Improving the Performance of the Mini 2000 Mass Spectrometer with a Triboelectric Nanogenerator Electrospray Ionization Source.
ACS Omega. 2018 Sep 28;3(9):12229-12234. doi: 10.1021/acsomega.8b01777. eCollection 2018 Sep 30.
5
Reinventing (Bio)chemical Analysis with Paper.
Anal Chem. 2018 Dec 4;90(23):13815-13825. doi: 10.1021/acs.analchem.8b04825. Epub 2018 Nov 19.
6
Implementation of a single quad MS detector in routine QC analysis of peptide drugs.
J Pharm Anal. 2016 Feb;6(1):24-31. doi: 10.1016/j.jpha.2015.09.002. Epub 2015 Sep 26.
7
Ambient Ionization and Miniature Mass Spectrometry Systems for Disease Diagnosis and Therapeutic Monitoring.
Theranostics. 2017 Jul 20;7(12):2968-2981. doi: 10.7150/thno.19410. eCollection 2017.
8
Direct Analysis of Organic Compounds in Liquid Using a Miniature Photoionization Ion Trap Mass Spectrometer with Pulsed Carrier-Gas Capillary Inlet.
J Am Soc Mass Spectrom. 2017 Aug;28(8):1702-1708. doi: 10.1007/s13361-017-1683-2. Epub 2017 Apr 21.
9
Reducing Space Charge Effects in a Linear Ion Trap by Rhombic Ion Excitation and Ejection.
J Am Soc Mass Spectrom. 2016 Jul;27(7):1256-62. doi: 10.1007/s13361-016-1393-1. Epub 2016 Apr 14.

本文引用的文献

1
Miniature mass spectrometers.
Annu Rev Anal Chem (Palo Alto Calif). 2009;2:187-214. doi: 10.1146/annurev-anchem-060908-155229.
3
Direct analysis of melamine in complex matrices using a handheld mass spectrometer.
Analyst. 2010 Apr;135(4):705-11. doi: 10.1039/b923427f. Epub 2009 Dec 24.
4
Ambient mass spectrometry.
Analyst. 2010 Apr;135(4):659-60. doi: 10.1039/c003812c. Epub 2010 Mar 8.
5
Development, characterization, and application of paper spray ionization.
Anal Chem. 2010 Mar 15;82(6):2463-71. doi: 10.1021/ac902854g.
9
Direct olive oil analysis by low-temperature plasma (LTP) ambient ionization mass spectrometry.
Rapid Commun Mass Spectrom. 2009 Oct;23(19):3057-62. doi: 10.1002/rcm.4220.
10
Low-temperature plasma probe for ambient desorption ionization.
Anal Chem. 2008 Dec 1;80(23):9097-104. doi: 10.1021/ac801641a.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验