Validation of co-registration of clinical lung ventilation and perfusion SPECT.

作者信息

Marsh Steven, Barnden Leighton, O'Keeffe Darin

机构信息

Department of Medical Physics and Bioengineering, Christchurch Hospital, Riccarton Avenue, Private Bag, 4710, Christchurch, New Zealand.

出版信息

Australas Phys Eng Sci Med. 2011 Apr;34(1):63-8. doi: 10.1007/s13246-011-0059-3. Epub 2011 Feb 18.

Abstract

Ventilation (V) and perfusion (Q) image data from 20 patients undergoing a routine clinical SPECT V/Q study were aligned via rigid coregistration using intrinsic image intensity values with the software package 'Qonsub'. Accuracy of the coregistration was quantified by an independent technique that used Tc-99m filled fiducial markers which had been adhered to the patient's skin prior to imaging. These were visible on both the V and Q scans, but were removed from the images for the coregistration step. The level of inter-marker displacement between scans was monitored to ensure it did not invalidate the assessment of coregistration accuracy. Once coregistered, results showed that for 65% of patients in the survey co-registration accuracy was better than 1 pixel (3.8 mm), 30% were co-registered with an accuracy of between 1 and 2 pixels and 5% were co-registered with an accuracy of between 2 and 3 pixels.