In order to explore the difference of intraocular pressure (IOP) at different points of cornea before and after laser in situ keratomileusis (LASIK), IOP was measured by Tono-Pen Tonometer at central cornea, pericentral cornea and limbus respectively and analyzed statistically. After LASIK, IOP was dropped significantly at central cornea and pericentral cornea (P<0.05), while no statistically significant change occurred at limbus (P>0.05). There was no statistically significant difference in IOP at different points before LASIK (F=0.110, P=0.896), but statistically significant difference was found after LASIK (F=7.375, P=0.001). It was suggested that reliable IOP after LASIK could be obtained from the limbus by Tono-Pen tonometer.
为探讨准分子原位角膜磨镶术(LASIK)前后角膜不同部位眼压(IOP)的差异,分别用非接触眼压计在角膜中央、角膜旁中央及角膜缘测量眼压并进行统计学分析。LASIK术后,角膜中央和角膜旁中央眼压显著下降(P<0.05),而角膜缘眼压无统计学意义的变化(P>0.05)。LASIK术前不同部位眼压无统计学差异(F=0.110,P=0.896),但术后有统计学差异(F=7.375,P=0.001)。提示非接触眼压计可通过测量角膜缘获得LASIK术后可靠的眼压值。