Comparison of intraocular pressures at different points in human's cornea before and after laser in situ keratomileusis with tono-pen tonometer.

作者信息

Li Xinyu, Li Guigang, Liu Lei, Li Jing

机构信息

Department of Ophthalmology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430030, China.

出版信息

J Huazhong Univ Sci Technolog Med Sci. 2011 Feb;31(1):128-130. doi: 10.1007/s11596-011-0163-3. Epub 2011 Feb 19.

Abstract

In order to explore the difference of intraocular pressure (IOP) at different points of cornea before and after laser in situ keratomileusis (LASIK), IOP was measured by Tono-Pen Tonometer at central cornea, pericentral cornea and limbus respectively and analyzed statistically. After LASIK, IOP was dropped significantly at central cornea and pericentral cornea (P<0.05), while no statistically significant change occurred at limbus (P>0.05). There was no statistically significant difference in IOP at different points before LASIK (F=0.110, P=0.896), but statistically significant difference was found after LASIK (F=7.375, P=0.001). It was suggested that reliable IOP after LASIK could be obtained from the limbus by Tono-Pen tonometer.

摘要

为探讨准分子原位角膜磨镶术(LASIK)前后角膜不同部位眼压(IOP)的差异,分别用非接触眼压计在角膜中央、角膜旁中央及角膜缘测量眼压并进行统计学分析。LASIK术后,角膜中央和角膜旁中央眼压显著下降(P<0.05),而角膜缘眼压无统计学意义的变化(P>0.05)。LASIK术前不同部位眼压无统计学差异(F=0.110,P=0.896),但术后有统计学差异(F=7.375,P=0.001)。提示非接触眼压计可通过测量角膜缘获得LASIK术后可靠的眼压值。