Takats Zoltan, Wiseman Justin M, Ifa Demian R, Cooks R Graham
CSH Protoc. 2008 Apr 1;2008:pdb.top37. doi: 10.1101/pdb.top37.
INTRODUCTIONDesorption electrospray ionization (DESI) is a desorption ionization (DI) method by nature, and, like matrix-assisted laser desorption/ionization (MALDI), it is used for the analysis of material present on a surface. DESI includes features reminiscent of electrospray ionization (ESI) in respect to both its instrumental and mechanistic aspects. However, the analyte in the DESI experiment is not in solution as in ESI. Instead, a microelectrospray ion source is used to produce charged droplets, ionic clusters, and/or gas-phase ions (depending on chosen experimental conditions), and these are directed at the sample surface. The sample is present in the ambient environment. An electrical potential of several kilovolts (kV) is applied to the spray solution, and pneumatic nebulization is used to assist in desolvation. Ionization of molecules present on the sample surface occurs upon the impact of the ESI-originated, charged particles with the surface. Surfaces include deposited samples on sample holder targets as well as surfaces of natural objects such as biological tissues or minerals.
引言
解吸电喷雾电离(DESI)本质上是一种解吸电离(DI)方法,与基质辅助激光解吸/电离(MALDI)一样,用于分析存在于表面的物质。DESI在仪器和机理方面都具有一些让人联想到电喷雾电离(ESI)的特征。然而,DESI实验中的分析物不像ESI那样处于溶液中。相反,使用微电喷雾离子源来产生带电液滴、离子簇和/或气相离子(取决于所选的实验条件),并将这些导向样品表面。样品存在于环境中。向喷雾溶液施加几千伏(kV)的电势,并使用气动雾化来辅助去溶剂化。当源自ESI的带电粒子撞击样品表面时,样品表面存在的分子发生电离。表面包括样品架靶上的沉积样品以及诸如生物组织或矿物等天然物体的表面。