Neuroimage. 2011 Jun 1;56(3):1244-50. doi: 10.1016/j.neuroimage.2011.02.080. Epub 2011 Mar 6.
Pseudocontinuous arterial spin labeling (PCASL), a newly proposed perfusion magnetic resonance imaging (MRI) technique, has the potential of providing a better balance between labeling efficiency and signal-to-noise ratio than conventional ASL methods. The advantage will not be exploitable until adequate reproducibility can be achieved. In this study, we investigated the reproducibility of PCASL on twelve healthy volunteers by taking into account the inclusion of correction for coil sensitivity (CS) and labeling efficiency (α). Each subject underwent two identical sessions scheduled two weeks apart. For comparison, pulsed ASL (PASL) imaging was also conducted using the flow-sensitive alternating inversion recovery method. MR imaging was performed on a 3T whole body system using a 12-channel phased array for reception. Blood flow was quantified for gray matter and assessed globally as well as in the flow territories of the internal carotid arteries and posterior cerebral arteries. Experimental results showed that CS accounted for 35% and 60% of inter-session variability in PCASL and PASL, respectively. With CS correction alone, coefficient of variance (CV) remained larger with PCASL than PASL (p=0.02). CV was 12% with CS-corrected PASL and 11% with PCASL when CS and α were corrected for in conjunction. We concluded that CS correction is necessary for ASL imaging when using phased array receiver and that after CS correction, PCASL requires α correction to provide reproducibility comparable to PASL.
伪连续动脉自旋标记(PCASL)是一种新提出的灌注磁共振成像(MRI)技术,与传统的 ASL 方法相比,它有可能在标记效率和信噪比之间取得更好的平衡。只有在能够达到足够的可重复性的情况下,才能利用这一优势。在这项研究中,我们通过纳入线圈灵敏度(CS)和标记效率(α)的校正,对 12 名健康志愿者的 PCASL 进行了可重复性研究。每个受试者进行两次完全相同的扫描,两次扫描间隔两周。为了进行比较,我们还使用流动敏感交替反转恢复方法进行了脉冲 ASL(PASL)成像。MR 成像在一台 3T 全身系统上进行,使用 12 通道相控阵接收器。对灰质的血流进行了量化,并在颈内动脉和大脑后动脉的血流区域进行了全局评估。实验结果表明,CS 分别占 PCASL 和 PASL 两次扫描之间变异性的 35%和 60%。单独进行 CS 校正后,PCASL 的 CV 仍然比 PASL 大(p=0.02)。在进行 CS 和 α 联合校正后,CS 校正的 PASL 的 CV 为 12%,PCASL 的 CV 为 11%。我们得出结论,当使用相控阵接收器进行 ASL 成像时,CS 校正是必要的,并且在 CS 校正后,PCASL 需要进行 α 校正,才能提供与 PASL 相当的可重复性。