Department of Radiology, Seoul National University Hospital, Jongno-gu, Seoul, Korea.
Acad Radiol. 2011 May;18(5):626-33. doi: 10.1016/j.acra.2010.12.013. Epub 2011 Mar 9.
The aims of this study were to compare accuracy between semiautomated and manual measurements of the longest diameter and volume of simulated hepatic tumors in phantoms and to evaluate the effects of slice thickness (ST) and reconstruction increment (RI) on accuracy.
Liver phantoms with 45 hypoattenuating and 45 hyperattenuating lesions of different sizes (diameter, 13.3-50.7 mm; volume, 0.4-54.0 mm(3)) and shapes (spherical or elliptical) were scanned using a 64-row multidetector computed tomographic scanner. Images were reconstructed with ST and RI settings of 0.75 and 0.7 mm, 1.0 and 0.7 mm, 1.5 and 1.0 mm, 3.0 and 2.0 mm, 3.0 and 3.0 mm, and 5.0 and 5.0 mm. The longest diameter and volume of each lesion were measured both manually and semiautomatically. To assess accuracy, measurements were compared to reference values by calculating absolute percentage error. Comparisons of absolute percentage error between methods and between ST and RI settings were performed using paired t tests. The degree of correlation between each measurement and a reference value was also assessed.
The semiautomated method showed significantly higher accuracy than the manual method in volume for most ST and RI settings (0.75 and 0.7 mm, 1.0 and 0.7 mm, and 1.5 and 1.0 mm in hypoattenuating lesions and all settings in hyperattenuating lesions; P < .05) and showed similar accuracy in diameter for all ST and RI settings regardless of lesion attenuation (P > .05). Semiautomated measurements also demonstrated higher correlation with reference values than the manual method for both diameter and volume. The absolute percentage error tended to be increased as ST and RI increased for both methods, and acceptable maximum ST and RI in semiautomated method were 1.5 and 1.0 mm.
Semiautomated computed tomographic measurement showed higher accuracy and correlation than the manual method in measuring the diameter and volume of hepatic lesions. The accuracy of both methods was highly dependent on z-axis resolution.
本研究旨在比较在体模中模拟肝肿瘤的最长直径和体积的半自动和手动测量的准确性,并评估层厚(ST)和重建增量(RI)对准确性的影响。
使用 64 排多层 CT 扫描仪对 45 个低衰减和 45 个高衰减的不同大小(直径 13.3-50.7mm;体积 0.4-54.0mm³)和形状(球形或椭圆形)的肝脏体模进行扫描。使用 ST 和 RI 设置为 0.75 和 0.7mm、1.0 和 0.7mm、1.5 和 1.0mm、3.0 和 2.0mm、3.0 和 3.0mm、5.0 和 5.0mm 对图像进行重建。手动和半自动测量每个病变的最长直径和体积。为了评估准确性,通过计算绝对百分比误差来比较测量值与参考值。使用配对 t 检验比较两种方法之间以及 ST 和 RI 设置之间的绝对百分比误差。还评估了每种测量值与参考值之间的相关性。
半自动方法在大多数 ST 和 RI 设置下(低衰减病变中的 0.75 和 0.7mm、1.0 和 0.7mm 以及 1.5 和 1.0mm,以及所有高衰减病变中的设置;P<0.05)在体积方面的准确性明显高于手动方法,在所有 ST 和 RI 设置下(无论病变衰减如何)在直径方面的准确性相似(P>0.05)。半自动测量值在直径和体积方面与参考值的相关性也高于手动方法。对于两种方法,随着 ST 和 RI 的增加,绝对百分比误差都有增加的趋势,半自动方法中可接受的最大 ST 和 RI 分别为 1.5mm 和 1.0mm。
半自动 CT 测量在测量肝病变的直径和体积方面比手动方法具有更高的准确性和相关性。两种方法的准确性都高度依赖于 z 轴分辨率。