Electrophoresis. 2011 Apr;32(8):884-9. doi: 10.1002/elps.201000420. Epub 2011 Mar 10.
The quantification of plasma lactate and evaluation of the lactate threshold by CE with capacitively coupled contactless conductivity is demonstrated. The only sample preparation needed was deproteinization with a ACN/methanol mixture. A solution of 10 mmol/L 2-morpholinoethanesulfonic acid monohydrate, 10 mmol/L DL-histidine, 70 μmol/L hexadecyltrimethylammonium bromide, pH 6.0 was found suitable as running buffer. Linearity was achieved for the concentration range of 10-1000 μmol/L with a correlation coefficient of 0.9994. The limit of detection (3 S/N) was determined as 3.2 μmol/L. Intra- and inter-day variabilities were less than 7% RSD. The suitability of the method could be demonstrated by analyzing various clinical samples, where the results correlated satisfactorily with those of an established enzymatic method.
通过电容耦合非接触式电导 CE 对血浆乳酸进行定量并评估乳酸阈,得到了验证。唯一需要的样品制备是用 ACN/甲醇混合物进行蛋白沉淀。发现 10 mmol/L 2-吗啉乙磺酸一水合物、10 mmol/L DL-组氨酸、70 μmol/L 十六烷基三甲基溴化铵、pH 6.0 的溶液适合作为运行缓冲液。在 10-1000 μmol/L 的浓度范围内实现了线性,相关系数为 0.9994。检测限(3 S/N)确定为 3.2 μmol/L。日内和日间变异系数均小于 7%RSD。该方法的适用性可通过分析各种临床样本得到证明,其结果与建立的酶法测定结果相关性良好。