The aim of this study was to establish reference concentrations of urinary strontium by inductively coupled plasma atomic emission spectrometry (ICP-AES).
For the determination of strontium, urine samples were collected from healthy Japanese (n=146; 115 males, 31 females; mean age, 33±9 years; age range, 18 to 58 years). The urine samples stored at or below -20°C were thawed with incubation at 40°C for 30 min and sediments were dissolved by vigorous shakings. Then, the samples were centrifuged at 3000 g for 5 min, and the supernatant was directly aspired into a P-5200-3600/1200 ICP-AES system from Hitachi Ltd., Tokyo, Japan.
A steeper increase in the S/N ratio and a good effective linearity of the calibration line was obtained at 407.771 nm in the range of 0-300 μg/L strontium standard solution. Urine samples having the same background signal as that of 18 MΩ cm ultrapure blank water, a good correspondence of the single peak pattern of the spectra, accuracy and precision of spike recovery were also confirmed. Urinary strontium concentrations showed a log-normal distribution and a geometric mean concentration of 143.9 μg/L, with 5-95% confidential interval of 40.9-505.8 μg/L.
The results of this study will be useful as guidelines for the biological monitoring of strontium in normal subjects and in individuals therapeutically or environmentally exposed to strontium.
本研究旨在通过电感耦合等离子体原子发射光谱法(ICP-AES)建立尿锶的参考浓度。
为了测定锶,收集了来自健康的日本成年人(n=146;115 名男性,31 名女性;平均年龄 33±9 岁;年龄范围 18 至 58 岁)的尿液样本。将储存在-20°C 或以下的尿液样本在 40°C 下孵育 30 分钟解冻,并用剧烈摇动溶解沉淀物。然后,将样品以 3000g 离心 5 分钟,将上清液直接吸入来自日本日立公司的 P-5200-3600/1200 ICP-AES 系统。
在 0-300μg/L 锶标准溶液范围内,在 407.771nm 处获得了更高的 S/N 比和更好的校准线有效线性度。尿液样本具有与 18MΩcm 超纯水相同的背景信号,光谱的单峰模式、准确度和精密度的良好对应性以及加标回收率也得到了确认。尿锶浓度呈对数正态分布,几何均数浓度为 143.9μg/L,5-95%置信区间为 40.9-505.8μg/L。
本研究的结果将有助于为正常人群和接受锶治疗或环境暴露的个体的锶生物监测提供指导。