Department of Radiation Oncology, William Beaumont Hospital, Royal Oak, MI, USA.
Int J Radiat Oncol Biol Phys. 2012 Apr 1;82(5):1636-41. doi: 10.1016/j.ijrobp.2011.02.011. Epub 2011 Apr 12.
Prolonged delivery times during daily cone-beam computed tomography (CBCT)-guided lung stereotactic body radiotherapy (SBRT) introduce concerns regarding intrafraction variation (IFV) of the mean target position (MTP). The purpose of this study was to evaluate the magnitude of the IFV-MTP and to assess target margins required to compensate for IFV and postonline CBCT correction residuals. Patient, treatment, and tumor characteristics were analyzed with respect to their impact on IFV-MTP.
A total of 126 patients with 140 tumors underwent 659 fractions of lung SBRT. Dose prescribed was 48 or 60 Gy in 12 Gy fractions. Translational target position correction of the MTP was performed via onboard CBCT. IFV-MTP was measured as the difference in MTP between the postcorrection CBCT and the posttreatment CBCT excluding residual error.
IFV-MTP was 0.2 ± 1.8 mm, 0.1 ± 1.9 mm, and 0.01 ± 1.5 mm in the craniocaudal, anteroposterior, and mediolateral dimensions and the IFV-MTP vector was 2.3 ± 2.1 mm. Treatment time and excursion were found to be significant predictors of IFV-MTP. An IFV-MTP vector greater than 2 and 5 mm was seen in 40.8% and 7.2% of fractions, respectively. IFV-MTP greater than 2 mm was seen in heavier patients with larger excursions and longer treatment times. Significant differences in IFV-MTP were seen between immobilization devices. The stereotactic frame immobilization device was found to be significantly less likely to have an IFV-MTP vector greater than 2 mm compared with the alpha cradle, BodyFIX, and hybrid immobilization devices.
Treatment time and respiratory excursion are significantly associated with IFV-MTP. Significant differences in IFV-MTP were found between immobilization devices. Target margins for IFV-MTP plus post-correction residuals are dependent on immobilization device with 5-mm uniform margins being acceptable for the frame immobilization device.
在日常锥形束计算机断层扫描(CBCT)引导下进行肺部立体定向体放射治疗(SBRT)时,延长的治疗时间会引起对平均靶区位置(MTP)的分次内变化(IFV)的关注。本研究的目的是评估 IFV-MTP 的幅度,并评估为补偿 IFV 和在线后 CBCT 校正残余所需的靶区边界。分析了患者、治疗和肿瘤特征对 IFV-MTP 的影响。
共 126 例 140 个肿瘤患者接受了 659 次肺部 SBRT 治疗。处方剂量为 48 或 60 Gy,分为 12 Gy 剂量。通过机载 CBCT 对 MTP 进行平移目标位置校正。IFV-MTP 定义为校正后 CBCT 与治疗后 CBCT 之间 MTP 的差值,不包括残余误差。
在颅尾、前后和左右方向上,IFV-MTP 分别为 0.2 ± 1.8mm、0.1 ± 1.9mm 和 0.01 ± 1.5mm,IFV-MTP 向量为 2.3 ± 2.1mm。治疗时间和位移被发现是 IFV-MTP 的显著预测因素。40.8%和 7.2%的治疗中分别出现了 IFV-MTP 向量大于 2mm 和 5mm 的情况。IFV-MTP 大于 2mm 的情况出现在体重较大、位移较大、治疗时间较长的患者中。在不同的固定装置之间,IFV-MTP 存在显著差异。与 alpha 摇篮、BodyFIX 和混合固定装置相比,立体定向框架固定装置的 IFV-MTP 向量大于 2mm 的可能性明显较低。
治疗时间和呼吸位移与 IFV-MTP 显著相关。在不同的固定装置之间,IFV-MTP 存在显著差异。对于 IFV-MTP 和校正后残余物,目标边界取决于固定装置,对于框架固定装置,5mm 的统一边界是可以接受的。