Takahashi Fumiaki, Sato Kaoru, Endo Akira, Ono Koji, Yoshitake Takayasu, Hasegawa Takayuki, Katsunuma Yasushi, Ban Nobuhiko, Kai Michiaki
Japan Atomic Energy Agency, 2-4 Shirakata, Tokai-mura, Ibaraki 319-1195, Japan.
Radiat Prot Dosimetry. 2011 Jul;146(1-3):241-3. doi: 10.1093/rpd/ncr160. Epub 2011 Apr 14.
A web system of WAZA-ARI is being developed to assess radiation dose to a patient in a computed tomography examination. WAZA-ARI uses one of organ dose data sets corresponding to the options selected by a user to describe examination conditions. The organ dose data have been derived by the Particle and Heavy Ion Transport code system, combined with Japanese male (JM) phantom. The configuration of JM phantom is adjusted to the averaged JM adult. In addition, a new phantom is introduced by removing arms from JM phantom to take into account for dose calculations in torso examinations. Some of the organ doses by JM phantom without arms are compared with results obtained by using a MIRD-type phantom, which was applied in some previous dosimetry systems.
正在开发一个WAZA - ARI网络系统,用于评估计算机断层扫描检查中患者所接受的辐射剂量。WAZA - ARI使用与用户选择的选项相对应的器官剂量数据集之一来描述检查条件。器官剂量数据是通过粒子与重离子传输代码系统,并结合日本男性(JM)体模得出的。JM体模的配置已根据成年日本男性的平均值进行了调整。此外,通过从JM体模中移除手臂引入了一种新体模,以考虑躯干检查中的剂量计算。将无手臂的JM体模得出的一些器官剂量与使用MIRD型体模获得的结果进行了比较,MIRD型体模曾应用于一些先前的剂量测定系统中。