Cranmer-Sargison Gavin, Kundapur Vijayananda, Tu Deluan, Ternes Shyanne, Vachhrajani Haresh, Sidhu Narinder P
Department of Medical Physics, Saskatchewan Cancer Agency, Canada.
Med Dosim. 2012 Summer;37(2):117-21. doi: 10.1016/j.meddos.2011.02.002. Epub 2011 Apr 15.
The goal of this work was to use daily kV-kV imaging and weekly cone-beam CT (CBCT) to evaluate rectal cancer patient position when treated on a new couch top belly board (BB). Quality assurance (QA) of the imaging system was conducted weekly to ensure proper performance. The positional uncertainty of the combined kV-kV image match and subsequent couch move was found to be no more than ± 1.0 mm. The average (1 SD) CBCT QA phantom match was anterior-posterior (AP) = -0.8 ± 0.2 mm, superior-inferior (SI) = 0.9 ± 0.2 mm, and left-right (LR) = -0.1 ± 0.1 mm. For treatment, a set of orthogonal kV-kV images were taken and a bony anatomy match performed online. Moves were made along each axis (AP, SI, and LR) and recorded for analysis. CBCT data were acquired once every 5 fractions for a total of 5 images per patient. The images were all taken after the couch move but before treatment. A 3-dimensional (3D-3D) bony anatomy auto-match was performed offline and the residual difference in position recorded for analysis. The average (± 1 SD) move required from skin marks, calculated over all 375 fractions (15 patients × 25 fractions/patient), were AP = -2.6 ± 3.7 mm, SI = -0.3 ± 4.9 mm, and LR = 1.8 ± 4.5 mm. The average residual difference in patient position calculated from the weekly CBCT data (75 total) were AP = -1.7 ± 0.4 mm, SI = 1.1 ± 0.6 mm, and LR = -0.5 ± 0.2 mm. These results show that the BB does provide simple patient positioning that is accurate to within ± 2.0 mm when using online orthogonal kV-kV image matching of the pelvic bony anatomy.
这项工作的目标是使用每日千伏-千伏成像和每周锥形束CT(CBCT)来评估直肠癌患者在新型床顶腹板(BB)上治疗时的体位。每周对成像系统进行质量保证(QA)以确保其正常运行。发现千伏-千伏图像匹配及随后的床移动的位置不确定性不超过±1.0毫米。CBCT QA体模匹配的平均值(1个标准差)为前后(AP)=-0.8±0.2毫米,上下(SI)=0.9±0.2毫米,左右(LR)=-0.1±0.1毫米。治疗时,采集一组正交千伏-千伏图像并在线进行骨骼解剖匹配。沿每个轴(AP、SI和LR)进行移动并记录以供分析。每5个分次采集一次CBCT数据,每位患者共采集5幅图像。所有图像均在床移动后但治疗前采集。离线进行三维(3D-3D)骨骼解剖自动匹配并记录位置的残余差异以供分析。在所有375个分次(15名患者×每位患者25个分次)上计算得出的从皮肤标记处所需的平均(±1个标准差)移动为AP=-2.6±3.7毫米,SI=-0.3±4.9毫米,LR=1.8±4.5毫米。根据每周CBCT数据(共75个)计算得出的患者位置的平均残余差异为AP=-1.7±0.4毫米,SI=1.1±0.6毫米,LR=-0.5±0.2毫米。这些结果表明,当使用骨盆骨骼解剖的在线正交千伏-千伏图像匹配时,腹板确实能提供精确到±2.0毫米以内的简单患者体位。