Department of Radiology, Taichung Veterans General Hospital, No. 160, Sec. 3, Taichung Harbor Road, Taichung 407, Taiwan.
Int J Cardiovasc Imaging. 2011 Jun;27(5):689-93. doi: 10.1007/s10554-011-9862-x. Epub 2011 Apr 19.
In computed tomography (CT), the term "artifact" is applied to any systematic discrepancy between the CT numbers in the reconstructed image and the true attenuation coefficients of the object. A partial ring artifact in cardiac CT has a unique pattern and has not been previously reported in real clinical practice. In this report, we will demonstrate the images and animations of a cardiac CT which is affected by a partial ring artifact due to a broken detector module. The different characteristics of ring artifacts on routine and cardiac CT, and the implications for clinical interpretation will also be described. In summary, when a partial ring artifact is noted incidentally on cardiac CT, we suggest interpreting the images by reviewing reconstructed images from different RR-intervals to avoid unnecessary repeated scans and additional radiation exposure.
在计算机断层扫描(CT)中,“伪影”一词用于描述重建图像中的 CT 值与物体真实衰减系数之间的任何系统性差异。心脏 CT 中的部分环状伪影具有独特的模式,在实际临床实践中尚未有报道。在本报告中,我们将展示由于探测器模块损坏而受到部分环状伪影影响的心脏 CT 的图像和动画。还将描述常规 CT 和心脏 CT 上环状伪影的不同特征及其对临床解释的影响。总之,当心脏 CT 上偶然出现部分环状伪影时,我们建议通过查看不同 RR 间期的重建图像来解释图像,以避免不必要的重复扫描和额外的辐射暴露。