Suppr超能文献

用于X射线光栅干涉测量的经验性束硬化和环形伪影校正(EBHC-GI)。

Empirical beam hardening and ring artifact correction for x-ray grating interferometry (EBHC-GI).

作者信息

Nelson Brandon J, Leng Shuai, Shanblatt Elisabeth R, McCollough Cynthia H, Koenig Thomas

机构信息

Graduate Program in Biomedical Engineering and Physiology, Mayo Clinic Graduate School of Biomedical Sciences, Rochester, MN, 55905, USA.

Department of Radiology, Mayo Clinic, Rochester, MN, 55905, USA.

出版信息

Med Phys. 2021 Mar;48(3):1327-1340. doi: 10.1002/mp.14672. Epub 2021 Jan 10.

Abstract

PURPOSE

Talbot-Lau grating interferometry enables the use of polychromatic x-ray sources, extending the range of potential applications amenable to phase contrast imaging. However, these sources introduce beam hardening effects not only from the samples but also from the gratings. As a result, grating inhomogeneities due to manufacturing imperfections can cause spectral nonuniformity artifacts when used with polychromatic sources. Consequently, the different energy dependencies of absorption, phase, and visibility contrasts impose challenges that so far have limited the achievable image quality. The purpose of this work was to develop and validate a correction strategy for grating-based x-ray imaging that accounts for beam hardening generated from both the imaged object and the gratings.

METHODS

The proposed two-variable polynomial expansion strategy was inspired by work performed to address beam hardening from a primary modulator. To account for the multicontrast nature of grating interferometry, this approach was extended to each contrast to obtain three sets of correction coefficients, which were determined empirically from a calibration scan. The method's feasibility was demonstrated using a tabletop Talbot-Lau grating interferometer micro-computed tomography (CT) system using CT acquisitions of a water sample and a silicon sample, representing low and high atomic number materials. Spectral artifacts such as cupping and ring artifacts were quantified using mean squared error (MSE) from the beam-hardening-free target image and standard deviation within a reconstructed image of the sample. Finally, the model developed using the water sample was applied to a fixated murine lung sample to demonstrate robustness for similar materials.

RESULTS

The water sample's absorption CT image was most impacted by spectral artifacts, but following correction to decrease ring artifacts, an 80% reduction in MSE and 57% reduction in standard deviation was observed. The silicon sample created severe artifacts in all contrasts, but following correction, MSE was reduced by 94% in absorption, 96% in phase, and 90% in visibility images. These improvements were due to the removal of ring artifacts for all contrasts and reduced cupping in absorption and phase images and reduced capping in visibility images. When the water calibration coefficients were applied to the lung sample, ring artifacts most prominent in the absorption contrast were eliminated.

CONCLUSIONS

The described method, which was developed to remove artifacts in absorption, phase, and normalized visibility micro-CT images due to beam hardening in the system gratings and imaged object, reduced the MSE by up to 96%. The method depends on calibrations that can be performed on any system and does not require detailed knowledge of the x-ray spectrum, detector energy response, grating attenuation properties and imperfections, or the geometry and composition of the imaged object.

摘要

目的

塔尔博特 - 劳光栅干涉测量法能够使用多色X射线源,扩展了适用于相衬成像的潜在应用范围。然而,这些源不仅会在样品中引入束硬化效应,也会在光栅中引入该效应。因此,由于制造缺陷导致的光栅不均匀性在与多色源一起使用时会引起光谱不均匀伪影。此外,吸收、相位和可见度对比度的不同能量依赖性带来了挑战,迄今为止限制了可实现的图像质量。这项工作的目的是开发并验证一种基于光栅的X射线成像校正策略,该策略考虑了成像物体和光栅产生的束硬化。

方法

所提出的双变量多项式展开策略的灵感来自于为解决来自初级调制器的束硬化所做的工作。为了考虑光栅干涉测量法的多对比度特性,该方法扩展到每个对比度以获得三组校正系数,这些系数通过校准扫描凭经验确定。使用台式塔尔博特 - 劳光栅干涉仪微型计算机断层扫描(CT)系统,通过对水样品和硅样品进行CT采集来证明该方法的可行性,水样品和硅样品分别代表低原子序数和高原子序数材料。使用与无束硬化目标图像的均方误差(MSE)以及样品重建图像内的标准偏差对诸如杯状伪影和环形伪影等光谱伪影进行量化。最后,将使用水样品开发的模型应用于固定的小鼠肺样品,以证明对类似材料的稳健性。

结果

水样品的吸收CT图像受光谱伪影影响最大,但在校正以减少环形伪影后,观察到MSE降低了80%,标准偏差降低了57%。硅样品在所有对比度中都产生了严重的伪影,但在校正后,吸收图像中的MSE降低了94%,相位图像中降低了96%,可见度图像中降低了90%。这些改进归因于所有对比度中环形伪影的消除以及吸收和相位图像中杯状伪影的减少以及可见度图像中盖帽伪影的减少。当将水校准系数应用于肺样品时,消除了吸收对比度中最突出的环形伪影。

结论

所描述的方法旨在消除由于系统光栅和成像物体中的束硬化而在吸收、相位和归一化可见度微型CT图像中产生的伪影,该方法使MSE降低了高达96%。该方法依赖于可以在任何系统上进行的校准,并且不需要详细了解X射线光谱、探测器能量响应、光栅衰减特性和缺陷,或者成像物体的几何形状和组成。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验