Fracture of the bearing surface is an infrequent cause of failure of a hip arthroplasty. Although well documented with ceramic heads, fracture of the metallic head is much rarer. We report a case of a fracture of the outer metallic head of a modular cemented bipolar hemiarthroplasty 2 years after the index procedure. Over time, the outer head lost its intended motion and assumed a vertical position. We hypothesized that this position caused asymmetrical loading with stress concentration at the poles, compounded by repeated impingement between the skirted inner cobalt-chromium (Cr-Co) head and the outer stainless steel head of this particular prosthesis. These were supported by the finite element studies. In addition, scanning electron microscopy and energy dispersive x-ray studies showed metallurgical defects that seemed to have initiated and/or accelerated the fracture. Although rare, this mode of failure calls for increased awareness, periodic follow-up, and quality control.
股骨头表面骨折是髋关节置换术失效的一个罕见原因。虽然陶瓷头已有充分的文献记载,但金属头骨折则更为罕见。我们报告了一例在指数手术后 2 年内发生的模块化粘结双极半髋关节置换术后外金属股骨头骨折的病例。随着时间的推移,外股骨头失去了预期的运动,并呈垂直位置。我们假设这种位置导致了不对称的负载,并在两极处产生了应力集中,加上裙边内钴铬(Cr-Co)头和该特定假体的外不锈钢头之间的反复撞击。这些都得到了有限元研究的支持。此外,扫描电子显微镜和能量色散 X 射线研究显示出似乎已经引发和/或加速了骨折的冶金缺陷。虽然罕见,但这种失效模式需要提高认识、定期随访和质量控制。