使用Gafchromic™ EBT2薄膜测量千伏级X射线束的反向散射因子。

The measurement of backscatter factors of kilovoltage X-ray beams using Gafchromic™ EBT2 film.

作者信息

Smith L, Hill R, Nakano M, Kim J, Kuncic Z

机构信息

Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia.

出版信息

Australas Phys Eng Sci Med. 2011 Jun;34(2):261-6. doi: 10.1007/s13246-011-0073-5. Epub 2011 Apr 29.

Abstract

Backscatter factors are essential in the determination of radiation dose for kilovoltage X-ray beams. The accurate measurement of backscatter factors in water (B (W)) is difficult and published values are based largely on Monte Carlo calculations. A number of studies have found that the measurement of B (W) in the energy range from 50 to 300 kVp is possible using Gafchromic EBT film, but this film is no longer commercially available. In this work, we evaluated whether the newer Gafchromic EBT2 film is suitable for the determination of B (W) for kilovoltage X-ray beams. B (W) were measured with Gafchromic EBT2 film for beam qualities of 50, 100 and 280 kVp and field sizes of 2, 3, 4 and 6 cm diameter and compared with tabulated values published in the AAPM TG-61 protocol. We found that rotation of the film position during readout had a small but non-negligible effect on the optical density readings. The agreement between measured and published B (W) was better than 3%, with the largest difference of occurring for the 2 cm diameter field with the 50 kVp X-ray beam. However, these differences are consistent with the total estimated uncertainty for the measurements, as calculated by the ISO GUM. Our results demonstrate that Gafchromic EBT2 film is a suitable dosimeter for B (W) measurements for clinical kilovoltage X-ray beams.

摘要

反向散射因子对于千伏级X射线束的辐射剂量测定至关重要。准确测量水中的反向散射因子(B(W))很困难,已发表的值很大程度上基于蒙特卡罗计算。多项研究发现,使用Gafchromic EBT胶片可以测量50至300 kVp能量范围内的B(W),但这种胶片已不再商业化供应。在这项工作中,我们评估了更新的Gafchromic EBT2胶片是否适用于测定千伏级X射线束的B(W)。使用Gafchromic EBT2胶片测量了50、100和280 kVp的射束质以及直径为2、3、4和6 cm的射野尺寸下的B(W),并与AAPM TG-61协议中公布的表格值进行了比较。我们发现,读出过程中胶片位置的旋转对光密度读数有微小但不可忽略的影响。测量值与公布的B(W)之间的一致性优于3%,最大差异出现在50 kVp X射线束的2 cm直径射野中。然而,这些差异与ISO GUM计算的测量总估计不确定度一致。我们的结果表明,Gafchromic EBT2胶片是用于临床千伏级X射线束B(W)测量的合适剂量计。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验