Department of Surgery, Klinikum rechts der Isar, Technische Universität München, München, Germany.
J Surg Res. 2012 Jun 15;175(2):191-8. doi: 10.1016/j.jss.2011.03.029. Epub 2011 Apr 13.
Technical progress in the surgical operating room (OR) increases constantly, facilitating the development of intelligent OR systems functioning as "safety backup" in the background of surgery. Precondition is comprehensive data retrieval to identify imminent risky situations and inaugurate adequate security mechanisms. Radio-frequency-identification (RFID) technology may have the potential to meet these demands.
We set up a pilot study investigating feasibility and appliance reliability of a stationary RFID system for real-time surgical sponge monitoring (passive tagged sponges, position monitoring: mayo-stand/abdominal situs/waste bucket) and OR team tracking (active transponders, position monitoring: right/left side of OR table).
In vitro: 20/20 sponges (100%) were detected on the mayo-stand and within the OR-phantom, however, real-time detection accuracy declined to 7/20 (33%) when the tags were moved simultaneously. All retained sponges were detected correctly. In vivo (animal): 7-10/10 sterilized sponges (70%-100%) were detected correctly within the abdominal cavity. OR-team: detection accuracy within the OR (surveillance antenna) and on both sides of the OR table (sector antenna) was 100%. Mean detection time for position change (left to right side and contrariwise) was 30-60 s. No transponder failure was noted.
This is the first combined RFID system that has been developed for stationary use in the surgical OR. Preclinical evaluation revealed a reliable sponge tracking and correct detection of retained textiles (passive RFID) but also demonstrated feasibility of comprehensive data acquisition of team motion (active RFID). However, detection accuracy needs to be further improved before implementation into the surgical OR.
手术室(OR)中的技术进步不断推进,这为智能 OR 系统的发展提供了便利,使其能够在手术过程中充当“安全备份”。前提是全面检索数据,以识别即将出现的危险情况,并启动足够的安全机制。射频识别(RFID)技术可能有潜力满足这些需求。
我们进行了一项试点研究,调查了用于实时手术海绵监测(被动标记海绵,位置监测:梅奥手术台/腹部位置/废物桶)和 OR 团队跟踪(主动应答器,位置监测:OR 手术台的右侧/左侧)的固定 RFID 系统的可行性和应用可靠性。
在体外:20/20 个(100%)海绵在梅奥手术台上和 OR 模拟体内被检测到,然而,当标签同时移动时,实时检测准确性下降到 7/20(33%)。所有保留的海绵都被正确检测到。在体内(动物):10 个/10 个(100%)已消毒的海绵在腹腔内被正确检测到。OR 团队:在 OR 内(监测天线)和 OR 手术台两侧(扇形天线)的检测准确率为 100%。位置变化(从左侧到右侧,反之亦然)的平均检测时间为 30-60 秒。未发现应答器故障。
这是第一个开发用于手术 OR 中固定使用的组合 RFID 系统。临床前评估显示,海绵跟踪可靠,并且可以正确检测到残留的纺织品(被动 RFID),同时还证明了全面获取团队运动数据的可行性(主动 RFID)。然而,在将其引入手术室之前,检测精度需要进一步提高。