Abstract

We have previously developed an online adaptive replanning technique to rapidly adapt the original plan according to daily CT. This paper reports the quality assurance (QA) developments in its clinical implementation for prostate cancer patients. A series of pre-clinical validation tests were carried out to verify the overall accuracy and consistency of the online replanning procedure. These tests include (a) phantom measurements of 22 individual patient adaptive plans to verify their accuracy and deliverability and (b) efficiency and applicability of the online replanning process. A four-step QA procedure was established to ensure the safe and accurate delivery of an adaptive plan, including (1) offline phantom measurement of the original plan, (2) online independent monitor unit (MU) calculation for a redundancy check, (3) online verification of plan-data transfer using an in-house software and (4) offline validation of actually delivered beam parameters. The pre-clinical validations demonstrate that the newly implemented online replanning technique is dosimetrically accurate and practically efficient. The four-step QA procedure is capable of identifying possible errors in the process of online adaptive radiotherapy and to ensure the safe and accurate delivery of the adaptive plans. Based on the success of this work, the online replanning technique has been used in the clinic to correct for interfractional changes during the prostate radiation therapy.

摘要

我们之前开发了一种在线自适应计划修正技术,可以根据每日 CT 迅速修正原始计划。本文报告了在前列腺癌患者的临床实施中,该技术在质量保证(QA)方面的进展。我们进行了一系列临床前验证测试,以验证在线计划修正过程的整体准确性和一致性。这些测试包括:(a)对 22 个个体化自适应计划进行体模测量,以验证其准确性和可交付性;(b)在线计划修正过程的效率和适用性。建立了四步 QA 程序,以确保自适应计划的安全准确交付,包括:(1)原始计划的离线体模测量;(2)在线独立监测器单位(MU)计算,以冗余检查;(3)使用内部软件在线验证计划数据传输;(4)实际传输光束参数的离线验证。临床前验证表明,新实现的在线计划修正技术在剂量学上是准确的,在实践中是高效的。四步 QA 程序能够识别在线自适应放疗过程中可能出现的错误,并确保自适应计划的安全准确交付。基于这项工作的成功,我们已将在线计划修正技术用于临床,以纠正前列腺放射治疗过程中的分次间变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验