Hisamoto H, Tsubuku M, Enomoto T, Watanabe K, Kawaguchi H, Koike Y, Suzuki K
Department of Applied Chemistry, Keio University, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama 223, Japan.
Anal Chem. 1996 Nov 1;68(21):3871-8. doi: 10.1021/ac951149+.
Rapid flow-through analysis (RFA) based on optode (optical chemical sensor) detection is proposed, and its performance is discussed using the RFA system equipped with a pH-sensitive optode as a model case. To demonstrate and understand the usefulness of the RFA system, long-lifetime pH-sensitive optodes were prepared using a hydrophilic poly(hydroxyethyl methacrylate) (poly-HEMA) membrane covalently immobilized with a dye containing an amino group such as Congo Red or Nile Blue. The response equation for the optode with the RFA system was proposed, and satisfactory estimation for rapid pH analysis was obtained. The proposed RFA can stand as an advanced method for conventional flow injection analysis.
提出了基于光极(光学化学传感器)检测的快速流通分析(RFA),并以配备pH敏感光极的RFA系统为例讨论了其性能。为了证明和理解RFA系统的实用性,使用共价固定有含氨基染料(如刚果红或尼罗蓝)的亲水性聚(甲基丙烯酸羟乙酯)(聚-HEMA)膜制备了长寿命pH敏感光极。提出了光极与RFA系统的响应方程,并获得了快速pH分析的满意估计。所提出的RFA可作为传统流动注射分析的一种先进方法。