Watson Sheri, Aguas Marita, Bienapfl Tracy, Colegrove Pat, Foisy Nancy, Jondahl Bonnie, Yosses Mary Beth, Yu Larissa, Anastas Zoe
Providence Portland Medical Center, Post Anesthesia Care Unit, Portland, OR 97213, USA.
J Perianesth Nurs. 2011 Jun;26(3):135-42. doi: 10.1016/j.jopan.2011.02.001.
The purpose of this study was to determine if blood pressure (BP) measured in the forearm or with an extra-long BP cuff in the upper arm accurately reflects BP measured in the upper arm with an appropriately sized BP cuff in patients with large upper arm circumference. A method-comparison design was used with a convenience sample of 49 PACU patients. Noninvasive blood pressures were obtained in two different locations (forearm; upper arm) and in the upper arm with an extra-long adult and recommended large adult cuff sizes. Data were analyzed by calculating bias and precision for the BP cuff size and location and Student's t-tests, with P < .0125 considered significant. Significantly higher forearm systolic (P < .0001) and diastolic (P < .0002) BP measurements were found compared to BP obtained in the upper arm with the reference standard BP cuff. Significantly higher systolic (t(48df) = 5.38, P < .0001), but not diastolic (t(48df) = 4.11, P < .019), BP differences were found for BP measured with the extra-long cuff at the upper arm site compared to the upper arm, reference standard BP. Findings suggest that the clinical practice of using the forearm or an extra-long cuff in the upper arm for BP measurement in post anesthesia patients with large upper arm circumferences may result in inaccurate BP values.
本研究的目的是确定在大上臂围患者中,在前臂测量的血压(BP)或使用上臂超长血压袖带测量的血压是否能准确反映使用合适尺寸血压袖带在上臂测量的血压。采用方法比较设计,选取了49名麻醉后护理单元(PACU)患者的便利样本。在两个不同位置(前臂;上臂)以及在上臂分别使用超长成人袖带和推荐的大成人袖带尺寸获取无创血压。通过计算血压袖带尺寸和位置的偏差与精密度以及进行学生t检验来分析数据,P < .0125被视为具有显著性。与使用标准参考血压袖带在上臂测量的血压相比,发现前臂收缩压(P < .0001)和舒张压(P < .0002)测量值显著更高。与上臂的标准参考血压相比,在上臂部位使用超长袖带测量的血压收缩压差异显著(t(48df) = 5.38, P < .0001),但舒张压差异不显著(t(48df) = 4.11, P < .019)。研究结果表明,对于上臂围较大的麻醉后患者,在前臂或上臂使用超长袖带进行血压测量的临床实践可能会导致血压值不准确。