在较宽的臂围范围内,单个刚性锥形袖口与标准尺寸膀胱耦合的自动振荡装置的准确性。

Accuracy of a single rigid conical cuff with standard-size bladder coupled to an automatic oscillometric device over a wide range of arm circumferences.

机构信息

Department of Clinical and Experimental Medicine, University of Padova, Padua, Italy.

出版信息

Hypertens Res. 2010 Nov;33(11):1186-91. doi: 10.1038/hr.2010.146. Epub 2010 Aug 5.

Abstract

Although the upper arm has the shape of a truncated cone, cylindrical cuffs and bladders are currently used for blood pressure (BP) measurement. The aims of this study were to describe upper arm characteristics and to test the accuracy of a standard adult-size conical cuff coupled to an oscillometric device over a wide range of arm circumferences. Arm characteristics were studied in 142 subjects with arm circumferences ranging from 22 to 45 cm (study 1). In a subset of 33 subjects with the same range of arm circumferences, a rigid conical cuff with standard-size bladder (12.6 × 24.0 cm) and a rigid cylindrical cuff (13.3 × 24.0 cm), both coupled to a Microlife BP A100 device, were tested according to the requirements of the protocol of the European Society of Hypertension (ESH; study 2). Study 1. In all subjects, upper-arm shape was tronco-conical with slant angles ranging from 89.5° to 82.2°. In a multiple linear regression analysis, only arm circumference was an independent predictor of conicity (P<0.001). Study 2. The rigid conical cuff passed all three phases of the ESH protocol for systolic and diastolic BPs. Mean device-observer BP differences obtained with the conical cuff were unrelated to arm circumference. When the rigid cylindrical cuff was used, ESH criteria were not satisfied, and the cuff overestimated systolic BPs in subjects with large arms. BP can be measured accurately with the use of a standard-size rigid conical cuff coupled to a BP A100 device for a wide range of arm circumferences.

摘要

尽管上臂呈截顶圆锥形,但目前仍使用圆柱形袖带和气囊来测量血压(BP)。本研究旨在描述上臂特征,并测试一种与示波法设备配套使用的标准成人尺寸锥形袖带在大范围臂围下的准确性。在臂围范围为 22 至 45cm 的 142 名受试者中(研究 1)研究了臂部特征。在 33 名具有相同臂围范围的受试者中,根据欧洲高血压学会(ESH)协议的要求,测试了具有标准尺寸气囊(12.6×24.0cm)的刚性锥形袖带和刚性圆柱形袖带(13.3×24.0cm)(研究 2),两者均与 Microlife BP A100 设备配套使用。研究 1. 在所有受试者中,上臂形状为截顶圆锥形,斜角范围为 89.5°至 82.2°。在多元线性回归分析中,只有臂围是锥形的独立预测因子(P<0.001)。研究 2. 刚性锥形袖带通过 ESH 协议的收缩压和舒张压的三个阶段。使用锥形袖带获得的设备-观察者血压差值与臂围无关。当使用刚性圆柱形袖带时,ESH 标准不满足,并且袖带高估了大臂围受试者的收缩压。使用与 BP A100 设备配套的标准尺寸刚性锥形袖带可以在大范围臂围内准确测量血压。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验