New design bioprosthesis: early outcome of Carpentier-Edwards PERIMOUNT Magna in the small annular aortic position.

作者信息

Tayama Eiki, Tomita Yukihiro, Takagi Kazuyoshi, Kosuga Tomokazu, Teshima Hideki, Imasaka Ken-ichi, Sonoda Hiromich

机构信息

Division of Cardiac Surgery, Clinical Research Center, Kyushu Medical Center, National Hospital Organization of Japan, 1-8-1 Jigyo-hama, Chuo-ku, 810-8563 Fukuoka, Japan.

出版信息

J Artif Organs. 2011 Dec;14(4):284-8. doi: 10.1007/s10047-011-0579-3. Epub 2011 Jun 21.

Abstract

The Carpentier-Edwards PERIMOUNT (CEP) Magna (Edwards Lifesciences, Irvine, CA, USA) is a newly developed bioprosthesis with an improved cuff design that allows its implantation into the smaller aortic annulus. We evaluated the hemodynamic performance of the CEP Magna for smaller aortic annulus cases. Patients who underwent aortic valve replacement for aortic stenosis receiving a Magna 19 mm (n = 13), were compared with a standard CEP (n = 19). In the 19-mm series, the real annular size was significantly smaller in the Magna than the standard (21.1 ± 0.8 vs. 19.8 ± 0.8 mm, p = 0.007). The Magna was significantly superior with respect to effective orifice area index (EOAI) at postoperative 3 months; however, no significant difference was seen in other factors (peak pressure gradient, left ventricular mass index, ejection fraction). At postoperative 3 months, despite the lack of statistical significance, the incidence of patient-prosthesis mismatch (PPM) was lower with the Magna. Using the same label size, the Magna can be implanted in a smaller aortic annulus with performance comparable with or better than hemodynamic performance with the standard CEP. The Magna is a useful prosthesis for the small aortic annular patient.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验