验证一种电磁定位设备与腹部 X 光片比较,以预测喂养管放置的准确性。

Verification of an electromagnetic placement device compared with abdominal radiograph to predict accuracy of feeding tube placement.

机构信息

St Vincent Hospital, Indianapolis, Indiana, USA.

出版信息

JPEN J Parenter Enteral Nutr. 2011 Jul;35(4):535-9. doi: 10.1177/0148607110387436.

Abstract

BACKGROUND

Use of an electromagnetic placement device (EMPD) facilitates placement of feeding tubes at the bedside. Standard practice for verification of feeding tube placement is via radiographic confirmation. The purpose of this research study was to assess the accuracy of placement of small-bore feeding tubes (SBFTs) as determined by EMPD interpretation compared with that of abdominal radiograph verification by a radiologist.

METHODS

This multicenter prospective study enrolled patients requiring bedside feeding tube placement. SBFTs were placed by an experienced investigator using the EMPD. Two abdominal radiographs were then obtained: one after initial SBFT placement and an additional radiograph after injection of contrast. Documentation of location based on clinician interpretation using the EMPD was then compared with radiologist interpretation.

RESULTS

The final sample size was 194 patients, including 18 pediatric patients. Patient age ranged from 12 days to 102 years. Median time for tube placement was 12 minutes. Of the 194 patients, only 1 patient had data showing discrepancies between the original EMPD verification and the final abdominal radiograph interpretation, providing a 99.5% agreement. No patient experienced complications during SBFT placement, and 15 patients had inadvertent airway placement that was avoided with the use of the EMPD.

CONCLUSIONS

There was a high percentage of agreement between EMPD and radiologic interpretation after contrast injection. The EMPD aided in avoiding inadvertent airway placement, with no patient complications. This device can be used safely at the bedside to facilitate placement of feeding tubes, leading to the delivery of early enteral nutrition.

摘要

背景

使用电磁定位设备(EMPD)可在床边方便地放置喂养管。验证喂养管放置位置的标准做法是通过放射照相确认。本研究的目的是评估 EMPD 解读确定的小口径喂养管(SBFT)放置的准确性,与放射科医生通过腹部 X 光片验证的准确性进行比较。

方法

这项多中心前瞻性研究纳入了需要床边喂养管放置的患者。由经验丰富的研究人员使用 EMPD 放置 SBFT。然后获得两张腹部 X 光片:一张是在初始 SBFT 放置后,另一张是在注射对比剂后。然后根据 EMPD 上的临床医生解读来记录位置,并与放射科医生的解读进行比较。

结果

最终样本量为 194 名患者,包括 18 名儿科患者。患者年龄从 12 天到 102 岁不等。置管中位时间为 12 分钟。在 194 名患者中,只有 1 名患者的原始 EMPD 验证与最终腹部 X 光片解读之间的数据存在差异,提供了 99.5%的一致性。在 SBFT 放置过程中,没有患者发生并发症,并且 15 名患者在使用 EMPD 时避免了意外的气道放置。

结论

在注射对比剂后,EMPD 和放射学解读之间存在高度一致性。EMPD 有助于避免意外的气道放置,没有患者出现并发症。该设备可安全地在床边使用,以方便喂养管的放置,从而尽早提供肠内营养。