Suppr超能文献

用于具有哑光反射背衬的放射变色胶片分析的扫描仪均匀性改进

Scanner uniformity improvements for radiochromic film analysis with matt reflectance backing.

作者信息

Butson Ethan, Alnawaf Hani, Yu Peter K N, Butson Martin

机构信息

The Illawarra Grammar School (TIGS), Western Ave, Mangerton, NSW 2500, Australia.

出版信息

Australas Phys Eng Sci Med. 2011 Sep;34(3):401-7. doi: 10.1007/s13246-011-0086-0. Epub 2011 Jul 7.

Abstract

A simple and reproducible method for increasing desktop scanner uniformity for the analysis of radiochromic films is presented. Scanner uniformity, especially in the non-scan direction, for transmission scanning is well known to be problematic for radiochromic film analysis and normally corrections need to be applied. These corrections are dependant on scanner coordinates and dose level applied which complicates dosimetry procedures. This study has highlighted that using reflectance scanning in combination with a matt, white backing material instead of the conventional gloss scanner finish, substantial increases in the scanner uniformity can be achieved within 90% of the scanning area. Uniformity within ±1% over the scanning area for our epsonV700 scanner tested was found. This is compared to within ±3% for reflection scanning with the gloss backing material and within ±4% for transmission scanning. The matt backing material used was simply 5 layers of standard quality white printing paper (80 g/m(2)). It was found that 5 layers was the optimal result for backing material however most of the improvements were seen with a minimum of 3 layers. Above 5 layers, no extra benefit was seen. This may eliminate the need to perform scanner corrections for position on the desktop scanners for radiochromic film dosimetry.

摘要

本文介绍了一种简单且可重复的方法,用于提高桌面扫描仪在分析放射变色胶片时的均匀性。众所周知,对于放射变色胶片分析而言,扫描仪的均匀性,尤其是在透射扫描的非扫描方向上,存在问题,通常需要进行校正。这些校正取决于扫描仪坐标和所施加的剂量水平,这使得剂量测定程序变得复杂。本研究强调,通过将反射扫描与哑光白色背衬材料相结合,而不是传统的光泽扫描仪表面处理,可以在扫描区域的90%范围内大幅提高扫描仪的均匀性。对于我们测试的爱普生V700扫描仪,在扫描区域内实现了±1%以内的均匀性。相比之下,使用光泽背衬材料进行反射扫描时均匀性在±3%以内,透射扫描时在±4%以内。所使用的哑光背衬材料仅仅是5层标准质量的白色打印纸(80 g/m²)。结果发现,5层是背衬材料的最佳效果,但至少3层时就可以看到大部分改进。超过5层则没有额外的益处。这可能消除了在桌面扫描仪上对放射变色胶片剂量测定进行扫描仪位置校正的必要性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验