North Coast Cancer Institute, Lismore, NSW, Australia.
Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia.
Phys Eng Sci Med. 2022 Sep;45(3):721-727. doi: 10.1007/s13246-022-01136-0. Epub 2022 May 30.
Radiochromic film is a good dosimeter choice for patient QA for complex treatment techniques because of its near tissue equivalency, high spatial resolution and established method of use. Commercial scanners are typically used for film dosimetry, with Epson scanners being the most common. Radiochromic film dosimetry is not straightforward having some well-defined problems which must be considered, one of the main ones being the Lateral Response Artefact (LRA) effect. Previous studies showed that the contributing factors to LRA are from the structure of the active ingredients of the film and the components and construction of the flatbed scanner. This study investigated the effect of the scanner lens on the LRA effect, as part of a wider investigation of scanner design effects and uncertainties. Gafchromic EBT3 films were irradiated with 40 × 40 cm field size 6 MV beams. Films were analysed using images captured by a Canon 7D camera utilising 18 mm, 50 mm and 100 mm focal length lenses compared to images scanned with a conventional Epson V700 scanner. The magnitude of the LRA was observed to be dependent on the focal length of the lens used to image the film. A substantial reduction in LRA was seen with the use of the 50 mm and 100 mm lenses, by factors of 3-5 for the 50 mm lens and 4-30 for the 100 mm lens compared to conventional desktop scanner techniques. This is expected to be from the longer focal length camera lens system being able to collect more light from distant areas compared to the scanner-based system. This provides an opportunity to design film dosimetry systems that minimise this artefact.
放射色胶片因其接近组织等效性、高空间分辨率和成熟的使用方法,是复杂治疗技术患者 QA 的理想剂量计选择。商用扫描仪通常用于胶片剂量测量,其中爱普生扫描仪最为常见。放射色胶片剂量测量并不简单,存在一些必须考虑的明确问题,其中主要问题之一是侧向响应伪影(LRA)效应。先前的研究表明,LRA 的影响因素来自胶片的活性成分的结构以及平板扫描仪的组件和结构。这项研究调查了扫描仪镜头对 LRA 效应的影响,这是对扫描仪设计效果和不确定性进行更广泛调查的一部分。Gafchromic EBT3 胶片用 40×40cm 射野大小 6MV 射线照射。使用佳能 7D 相机拍摄的图像对胶片进行分析,相机分别使用了 18mm、50mm 和 100mm 焦距镜头,与传统爱普生 V700 扫描仪扫描的图像进行了比较。观察到 LRA 的大小取决于用于成像胶片的镜头焦距。与传统台式扫描仪技术相比,使用 50mm 和 100mm 镜头可显著减少 LRA,50mm 镜头的减少幅度为 3-5 倍,100mm 镜头的减少幅度为 4-30 倍。这预计是由于长焦距相机镜头系统能够比基于扫描仪的系统从更远的区域收集更多的光。这为设计最小化这种伪影的胶片剂量测量系统提供了机会。