使用标记药丸通过磁共振成像(MRI)对颅内病变进行术前定位

Preoperative localization of intracranial lesions with MRI using marking pills.

作者信息

Shibata Sumiya, Kunieda Takeharu, Adachi Hidemitsu, Ueno Yasushi, Kohara Nobuo, Sakai Nobuyuki

机构信息

Department of Neurosurgery, Kobe City Medical Center General Hospital, Japan.

出版信息

Clin Neurol Neurosurg. 2011 Dec;113(10):854-8. doi: 10.1016/j.clineuro.2011.06.004. Epub 2011 Jul 20.

Abstract

OBJECTIVE

To describe a simple technique for preoperative surface localization of intracranial lesions.

METHODS

11 pills in total, including Alfarol (alfacalcidol) capsules, were affixed to a phantom with adhesive tape and a MRI scan was performed. The visibility of the pills and any spatial errors in determining their locations were evaluated. Between June 2006 and April 2009, we employed Alfarol capsules as a skin marker in MRI in clinical surgical cases.

RESULTS

Alfarol capsules, whose actual size is 5.6 mm in diameter, were identified as a hyperintense spot at a size of 4.2, 4.2, and 4.5mm in diameter in T1-weighted, T2-weighted, and FLAIR (fluid attenuated inversion recovery) sequence images, respectively. The size discrepancies were within 1.4 mm. The average spatial errors were 0.7, 0.6, and 0.7 mm in T1-weighted, T2-weighted, and FLAIR sequence images, respectively. Other pills were not identified in the MRI scans. During this 35-month period, 8 patients underwent preoperative MRI-guided localization at our institution. There were 5 men and 3 women in whom 8 biopsies were performed. In all cases, the result of the biopsy was positive and useful for the treatment that followed. No perioperative complications were encountered.

CONCLUSION

Alfarol capsule can be used as an external skin marker. Our simple and inexpensive method is a useful addition to preoperative evaluation of superficial intracranial lesions.

摘要

目的

描述一种用于颅内病变术前体表定位的简单技术。

方法

总共11粒药丸,包括阿法骨化醇胶囊,用胶带粘贴在模拟体模上并进行磁共振成像(MRI)扫描。评估药丸的可视性以及确定其位置时的任何空间误差。在2006年6月至2009年4月期间,我们在临床手术病例中使用阿法骨化醇胶囊作为MRI中的皮肤标记物。

结果

阿法骨化醇胶囊实际直径为5.6毫米,在T1加权、T2加权和液体衰减反转恢复(FLAIR)序列图像中分别被识别为直径为4.2、4.2和4.5毫米的高强度斑点。尺寸差异在1.4毫米以内。在T1加权、T2加权和FLAIR序列图像中的平均空间误差分别为0.7、0.6和0.7毫米。在MRI扫描中未识别出其他药丸。在这35个月期间,8例患者在我们机构接受了术前MRI引导下的定位。有5名男性和3名女性,共进行了8次活检。在所有病例中,活检结果均为阳性,对后续治疗有用。未遇到围手术期并发症。

结论

阿法骨化醇胶囊可作为外部皮肤标记物。我们简单且廉价的方法是对颅内浅表病变术前评估的有益补充。